# epoc Blood Analysis System

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/epoc-blood-analysis-system-chl>
- Manufacturer: [Epocal Inc](https://devicerts.com/manufacturers/epocal-inc) (Canada)
- Device class: Class II (US FDA)
- FDA product code: CHL

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K200107 · decision 2020-06-05 · decision SESE · status: Active

## Recalls

- Z-2005-2026 · Class II · ongoing · initiated 2026-04-02: Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.
- Z-2272-2025 · Class II · ongoing · initiated 2025-07-07 · matched through shared submission K200107, may concern other models: Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L. The maximum bias observed was -14 mmol/L, which occurred at a higher sodium concentration around 150 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing.
- Z-1685-2025 · Class II · ongoing · initiated 2025-03-19: The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
- Z-1684-2025 · Class II · ongoing · initiated 2025-03-19: The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
- Z-0741-2025 · Class II · ongoing · initiated 2024-11-22: Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
- Z-0735-2025 · Class II · ongoing · initiated 2024-11-22: Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
- Z-0739-2025 · Class II · ongoing · initiated 2024-11-22: Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
- Z-0740-2025 · Class II · ongoing · initiated 2024-11-22: Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
- Z-0734-2025 · Class II · ongoing · initiated 2024-11-22: Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
- Z-0736-2025 · Class II · ongoing · initiated 2024-11-22: Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
- Z-0738-2025 · Class II · ongoing · initiated 2024-11-22: Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
- Z-0737-2025 · Class II · ongoing · initiated 2024-11-22: Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
- Z-0286-2025 · Class II · ongoing · initiated 2024-09-24: Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. One customer complaint has been associated with this issue. No adverse events have been reported. All affected lots have recently expired before the issuance of the recall.
- Z-0601-2024 · Class II · ongoing · initiated 2023-10-19: Siemens Healthcare Diagnostics Inc. has confirmed a potential issue that could occur under certain conditions, a software problem is causing erroneous results to be printed. When the results are run on the epoc NXS Host and Delete Blood Tests feature is enabled, the results that are generated at the time of testing are correct; however, when this issue occurs, results that are printed, saved and/or transferred to a data manager, may include unselected analytes that are from a previous patient or QC test. The worst possible outcome may result in a failure to diagnose hyperkalemia, due to an erroneous depressed potassium level or inappropriate treatment to hyperglycemia due to erroneous glucose reading.
- Z-2477-2023 · Class II · ongoing · initiated 2023-06-01: There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
- Z-2476-2023 · Class II · ongoing · initiated 2023-06-01: There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
- Z-2478-2023 · Class II · ongoing · initiated 2023-06-01: There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
- Z-1215-2023 · Class II · ongoing · initiated 2022-12-12: pO2 High Recovery in Quality Control When Adjusted for Barometric Pressure
- Z-2970-2020 · Class II · terminated · initiated 2020-08-28 · matched through shared submission K200107, may concern other models: Sporadically inconsistent discrepant (low bias) glucose results on card lot 01-20095-10. Potential exists for a delay in diagnosis of hyperglycemia or unnecessary treatment for hypoglycemia.

## Identifiers

- 510(k) Number: K200107 [FDA 510(k)]
- FDA product code: CHL [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 862.1120 [FDA 510(k)]
- Advisory committee: Clinical Chemistry [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: epoc Blood Analysis System, Epocal Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/epoc-blood-analysis-system-chl (data as of 2026-11-18).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-11
- Statuses computed as of: 2026-11-18
- JSON: <https://devicerts.com/devices/epoc-blood-analysis-system-chl.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/epoc-blood-analysis-system-chl
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.