  "format": "devicerts.device.v1",
  "cite_as": "Equinoxe 301-09-56, Exactech, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/equinoxe-301-09-56 (data as of 2026-11-22).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/equinoxe-301-09-56",
    "json_url": "https://devicerts.com/devices/equinoxe-301-09-56.json",
    "markdown_url": "https://devicerts.com/devices/equinoxe-301-09-56.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA GUDID (AccessGUDID)",
        "url": "https://accessgudid.nlm.nih.gov/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-23",
    "status_as_of": "2026-11-22",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/equinoxe-301-09-56",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Equinoxe 301-09-56",
    "registry_name": "Equinoxe 301-09-56",
    "url": "https://devicerts.com/devices/equinoxe-301-09-56",
    "manufacturer": {
      "name": "Exactech, Inc.",
      "url": "https://devicerts.com/manufacturers/exactech-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "Catalogue number",
        "values": [
          {
            "value": "301-09-56",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "Primary DI / UDI-DI",
        "values": [
          {
            "value": "10885862559944",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K042021",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          },
          {
            "value": "K063569",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          },
          {
            "value": "K082702",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          },
          {
            "value": "K102951",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          },
          {
            "value": "K121067",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          },
          {
            "value": "K140063",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          },
          {
            "value": "K162726",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "PHX",
      "KWS",
      "KWT",
      "HSD"
    ],
    "emdn": [],
    "gmdn": [
      {
        "code": "45114",
        "term": "Bone-resection orthopaedic reamer, reusable"
      }
    ],
    "description": null,
    "noindex": true
  },
  "authorizations": [
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "GUDID listing",
      "number": "10885862559944",
      "number_label": "UDI-DI",
      "device_udi_di": null,
      "date": "2022-07-28",
      "date_kind": "listed",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Listed",
      "status_note": "GUDID: in commercial distribution",
      "derived_from_gudid": false
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K162726",
      "number_label": null,
      "device_udi_di": null,
      "date": "2017-02-01",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K140063",
      "number_label": null,
      "device_udi_di": null,
      "date": "2014-02-11",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K121067",
      "number_label": null,
      "device_udi_di": null,
      "date": "2012-12-06",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K102951",
      "number_label": null,
      "device_udi_di": null,
      "date": "2010-11-04",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K082702",
      "number_label": null,
      "device_udi_di": null,
      "date": "2008-10-10",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K063569",
      "number_label": null,
      "device_udi_di": null,
      "date": "2007-02-23",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K042021",
      "number_label": null,
      "device_udi_di": null,
      "date": "2004-10-26",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0537-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-09-05",
      "reason": "Due to products that have not completed design updates and performance testing being released/distributed.  Products may not conform to design and performance criteria.",
      "applies": "this device",
      "via": "K042021"
    },
    {
      "number": "Z-0536-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-09-05",
      "reason": "Due to products that have not completed design updates and performance testing being released/distributed.  Products may not conform to design and performance criteria.",
      "applies": "this device",
      "via": "K042021"
    },
    {
      "number": "Z-0535-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-09-05",
      "reason": "Due to products that have not completed design updates and performance testing being released/distributed.  Products may not conform to design and performance criteria.",
      "applies": "this device",
      "via": "K042021"
    },
    {
      "number": "Z-2561-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-08-15",
      "reason": "Reverse Shoulder humeral liners have an articular surface position outside of the dimensional specification within the labeling.",
      "applies": "this device",
      "via": "K063569"
    },
    {
      "number": "Z-1419-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-03-06",
      "reason": "The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).",
      "applies": "this device",
      "via": "K063569"
    },
    {
      "number": "Z-1405-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-03-06",
      "reason": "The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).",
      "applies": "this device",
      "via": "K042021"
    },
    {
      "number": "Z-1422-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-03-06",
      "reason": "The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).",
      "applies": "this device",
      "via": "K063569"
    },
    {
      "number": "Z-1424-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-03-06",
      "reason": "The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).",
      "applies": "this device",
      "via": "K063569"
    },
    {
      "number": "Z-1403-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-03-06",
      "reason": "The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).",
      "applies": "this device",
      "via": "K042021"
    },
    {
      "number": "Z-1418-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-03-06",
      "reason": "The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).",
      "applies": "this device",
      "via": "K063569"
    },
    {
      "number": "Z-1404-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-03-06",
      "reason": "The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).",
      "applies": "this device",
      "via": "K042021"
    },
    {
      "number": "Z-1423-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-03-06",
      "reason": "The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).",
      "applies": "this device",
      "via": "K063569"
    },
    {
      "number": "Z-1402-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-03-06",
      "reason": "The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).",
      "applies": "this device",
      "via": "K042021"
    },
    {
      "number": "Z-1425-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-03-06",
      "reason": "The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).",
      "applies": "this device",
      "via": "K063569"
    },
    {
      "number": "Z-0579-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-11-21",
      "reason": "Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.",
      "applies": "this device",
      "via": "K063569"
    },
    {
      "number": "Z-0884-2021",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2020-11-09",
      "reason": "Potential for incorrect Instructions For Use (IFU) included with an Equinoxe Humeral Stem, Primary, Press Fit, 11mm package.",
      "applies": "this device",
      "via": "K042021"
    },
    {
      "number": "Z-3020-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-08-24",
      "reason": "Exactech is recalling the Equinoxe Reverse Shoulder Compression Screw/Locking Cap Kit, because it may contain a incorrect length screw.",
      "applies": "this device",
      "via": "K063569"
    },
    {
      "number": "Z-3019-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-08-24",
      "reason": "Exactech is recalling the Equinoxe Reverse Shoulder Compression Screw/Locking Cap Kit, because it may contain a incorrect length screw.",
      "applies": "this device",
      "via": "K063569"
    },
    {
      "number": "Z-0958-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-12-05",
      "reason": "Possibility that the Equinoxe Preserve Humeral  Stem, 8mm, was inadvertently packaged with an Optetrak Knee System instructions for use (IFU) rather than the  Equinoxe Preserve Humeral Stem IFU as intended .",
      "applies": "this device",
      "via": "K162726"
    },
    {
      "number": "Z-0012-2019",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2018-08-21",
      "reason": "The Equinoxe Reverse Drill Bit Kits may contain two 2mm diameter drill bits instead of one 2mm diameter drill bit and one 3.2mm diameter drill bit as intended.",
      "applies": "this device",
      "via": "K063569"
    },
    {
      "number": "Z-0900-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-10-25",
      "reason": "Disengaged screw head could potentially not be retained in the Torque Defining Screwdriver.",
      "applies": "this device",
      "via": "K063569"
    },
    {
      "number": "Z-1459-2015",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-02-13",
      "reason": "Incorrectly packaged.  Outer and inner labeling may incorrectly identify the size of the enclosed device.",
      "applies": "this device",
      "via": "K042021"
    },
    {
      "number": "Z-2115-2012",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-06-15",
      "reason": "Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year",
      "applies": "this device",
      "via": "K042021"
    },
    {
      "number": "Z-2117-2012",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-06-15",
      "reason": "Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year",
      "applies": "this device",
      "via": "K063569"
    },
    {
      "number": "Z-2119-2012",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-06-15",
      "reason": "Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year",
      "applies": "this device",
      "via": "K102951"
    },
    {
      "number": "Z-2116-2012",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-06-15",
      "reason": "Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year",
      "applies": "this device",
      "via": "K102951"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identity",
      "fields": [
        {
          "label": "Trade name",
          "values": [
            {
              "value": "Equinoxe",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Generic name",
          "values": [
            {
              "value": "Bone-resection orthopaedic reamer, reusable",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Model / Ref / Catalog",
          "values": [
            {
              "value": "301-09-56",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "Primary DI / UDI-DI",
          "values": [
            {
              "value": "10885862559944",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K042021",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "K063569",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "K082702",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "K102951",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "K121067",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "K140063",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "K162726",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "PHX",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "KWS",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "KWT",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "HSD",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "GMDN term",
          "values": [
            {
              "value": "Bone-resection orthopaedic reamer, reusable",
              "code": "45114",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Regulation number",
          "values": [
            {
              "value": "888.3660",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "888.3650",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "888.3690",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "GUDID distribution status",
          "values": [
            {
              "value": "In Commercial Distribution",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Clinical & Handling",
      "fields": [
        {
          "label": "Sterilisation method",
          "values": [
            {
              "value": "Moist Heat or Steam Sterilization",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Properties",
      "fields": [
        {
          "label": "Sterilise before use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Prescription use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Lot / batch number",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Supplied sterile",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Single use",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Implantable",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Contains latex",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Over-the-counter",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Convenience kit",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Combination product",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "HCT/P",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Premarket exempt",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Serial number",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Expiration date",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Manufacturing date",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    }
  ]
}