Equinoxe 304-21-09
Equinoxe 304-21-09 is a device of GMDN type partial shoulder prosthesis (FDA Class II) made by Exactech, Inc., cleared under FDA 510(k) K101909 on 2 Aug 2010 (1 earlier submission) and listed in FDA GUDID since 30 Oct 2015 (UDI-DI 10885862083357), active as of 23 Oct 2026.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10885862083357 | Date30 Oct 2015 | StatusListed |
| United States | FDA 510(k)SESE | NumberK101909 | Date2 Aug 2010 | StatusActive |
Official records
- AccessGUDID record 10885862083357
- FDA 510(k) database K101909
- FDA 510(k) database K092900
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K101909 decided 2 Aug 2010 Active
- FDA GUDID UDI-DI 10885862083357 listed 30 Oct 2015 Listed
Recalls
-
- Terminated Z-0307-2021
- Terminated Z-0308-2021
The device may be incorrectly laser marked as a right stem, when it should have been laser marked as a left stem.
-
- Terminated Z-0872-2020
- Terminated Z-0873-2020
Labels with incorrect translations were identified by Exactech is a distributor in the Netherlands. The translations on the outer labels include the word " long" in all languages other than the original English version. The implants are standard length fracture stems.
Timeline
- Cleared FDA 510(k) K092900
- Cleared FDA 510(k) K101909
- Listed FDA GUDID 10885862083357
- Recall initiated FDA recalls Z-0872-2020
- Recall initiated FDA recalls Z-0873-2020
- Recall initiated FDA recalls Z-0307-2021
- Recall initiated FDA recalls Z-0308-2021
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Equinoxe
- Generic name
- Partial shoulder prosthesis
- Model / Ref / Catalog
- 304-21-09
- Primary DI / UDI-DI
- 10885862083357
- FDA product code
- KWS, KWT
- GMDN term
- Partial shoulder prosthesis
- Device class
- Class II
- Regulation number
- 888.3660, 888.3650
- GUDID distribution status
- In Commercial Distribution
- Size
- Outer Diameter — 8.5 Millimeter
Similar devices
Other devices filed under FDA product code KWS (Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented).
- Anatomic Shoulder Prosthesis 04.01.0092 by Medacta International SA
- ENCORE SHOULDER PRIMARY HUMERAL STEM, MODEL 500-00, ENCORE SHOULDER HUMERAL HEAD, MODEL 520-38 by ENCORE MEDICAL, L.P.
- SOLAR REUNION FRACTURE STEM by Howmedica Osteonics Corp.
- Alliance® SBGL6031 by Zimmer, Inc.
- Anatomical Shoulder™ Domelock® 01.04212.565 by Zimmer, Inc.
- Anatomic Shoulder Prosthesis 04.01.0039 by Medacta International SA
- BLUEPRINT™ PATIENT SPECIFIC INSTRUMENTATION MWJ004 by TORNIER, INC.
- TORNIER PERFORM™ REVERSED MWJ10461 by TORNIER, INC.
- VAPORIZED HYDROGEN PEROXIDE (VHP) STERILIZATION OF UNMWPE COMPONENTS IN THE COFIELD TOTAL SHOULDER SYSTEM, THE NEER II by Smith & Nephew, Inc.
- LINK Embrace Shoulder System 642-010/04 by WALDEMAR LINK GmbH & Co. KG