Skip to content

Equinoxe Reverse Shoulder System Fracture Humeral Adapter Tray and Locking Screw

by Exactech, Inc.

510(k) K073688

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

1 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK073688 Date29 Feb 2008ClearedStatusActive

Recalls

1 recall · FDA

    • Terminated Z-2118-2012 FDA Class II initiated 15 Jun 2012 on shared 510(k) K073688

    Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
PHX
Device Class
Class II
Regulation
21 CFR Part 888.3660
Advisory Committee
Orthopedic
FDA Decision
SESE Substantially Equivalent

Other devices from Exactech, Inc.

See all devices from Exactech, Inc. →