Equinoxe Reverse Shoulder System
510(k) K063569
Equinoxe Reverse Shoulder System is a Class II device made by Exactech, Inc., cleared under FDA 510(k) K063569 on 23 Feb 2007, active as of 22 Oct 2026. 9 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
9 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK063569 | Date23 Feb 2007 | StatusActive |
Official records
- FDA 510(k) database K063569
Recalls
-
- Ongoing Z-2561-2025
Reverse Shoulder humeral liners have an articular surface position outside of the dimensional specification within the labeling.
-
- Ongoing Z-1425-2024
- Ongoing Z-1423-2024
- Ongoing Z-1419-2024
- Ongoing Z-1422-2024
- Ongoing Z-1424-2024
- Ongoing Z-1418-2024
The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).
-
- Ongoing Z-0579-2024
Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.
-
- Ongoing Z-0012-2019
The Equinoxe Reverse Drill Bit Kits may contain two 2mm diameter drill bits instead of one 2mm diameter drill bit and one 3.2mm diameter drill bit as intended.
-
- Terminated Z-3019-2020
- Terminated Z-3020-2020
Exactech is recalling the Equinoxe Reverse Shoulder Compression Screw/Locking Cap Kit, because it may contain a incorrect length screw.
-
- Terminated Z-0900-2017
Disengaged screw head could potentially not be retained in the Torque Defining Screwdriver.
-
- Terminated Z-2117-2012
Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year
Timeline
- Recall initiated FDA recalls Z-3019-2020
- Recall initiated FDA recalls Z-3020-2020
- Recall initiated FDA recalls Z-0579-2024
- Recall initiated FDA recalls Z-1425-2024
- Recall initiated FDA recalls Z-1423-2024
- Recall initiated FDA recalls Z-1419-2024
- Recall initiated FDA recalls Z-1422-2024
- Recall initiated FDA recalls Z-1424-2024
- Recall initiated FDA recalls Z-1418-2024
- Recall initiated FDA recalls Z-2561-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- PHX
- Device class
- Class II
- Regulation
- 21 CFR Part 888.3660
- Advisory committee
- Orthopedic
- FDA decision
- SESE Substantially Equivalent