Esprit Ventilator
510(k) K981072
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)ST | NumberK981072 | Date12 Nov 1998 | StatusActive |
Official records
- FDA 510(k) database K981072
From the FDA record
- Received by the FDA
- 24 Mar 1998
- FDA decision
- 12 Nov 1998 (Substantially Equivalent - Subject to Tracking Reg.)
- FDA review time
- 233 days from receipt to decision
Recalls
-
- Terminated Z-0485-2015
If the power supply fan mounting screws are installed incorrectly, the ends of the screws may touch the power supply printed circuit board assembly (PCBA) and short to ground. This may prevent the ventilator to switch to battery power if AC power is disrupted.
-
- Terminated Z-0008-2015
A failure of a specific component may prevent the ventilators from operating on AC power or transitioning back to AC power when operating on battery. Additionally, if a battery is not present or is depleted the ventilators will not operate. This failure will cause the loss of ventilator support, which may result in hypercarbia or hypoxemia.
-
- Terminated Z-1813-2009
The recall was initiated because through routine product monitoring Respironics discovered that under certain conditions, when operating on AC line power, a portion of the current from the External Battery charger is erroneously diverted to the ventilator. When this occurs, the External Battery charger may not transition correctly. Prolonged operation in this condition results in overcharging and
-
- Terminated Z-2003-2008
- Terminated Z-2004-2008
- Terminated Z-2002-2008
This action is being taken to address power supply failures on the Esprit Ventilator which have occurred in some units. These failures, which were discovered through routine product monitoring, have been reported in countries that utilize an input voltage of greater than 200 volts AC. Some customers have reported a burning smell coming from the ventilator.
-
- Terminated Z-2084-2013
The power to the Esprit Display's Backlight is interrupted, causing the GUI to flicker or go dark and making the display unreadable.
-
- Terminated Z-0047-2007
This action is being initiated due to power supply snubber board failures on certain Esprit Ventilators in countries that utilize operating voltages of 200-240 Volts AC and will involve the replacement of the Esprit Ventilator power supply snubber board. Returned power supplies have shown evidence of overheating and in some cases in which a fan also failed, have shown signs of fire damage.
-
- Terminated Z-1230-04
Flow sensors would exhibit drift.
-
- Terminated Z-0951-03
Material hardness on original design check valve may cause premature failure.
-
- Terminated Z-0903-04
Ventilator would spontaneously suspend ventilatory support to patients due to a hardware/software compatibility issue.
Timeline
- Recall initiated FDA recalls Z-0951-03
- Recall initiated FDA recalls Z-1230-04
- Recall initiated FDA recalls Z-0047-2007
- Recall initiated FDA recalls Z-2084-2013
- Recall initiated FDA recalls Z-2003-2008
- Recall initiated FDA recalls Z-2004-2008
- Recall initiated FDA recalls Z-2002-2008
- Recall initiated FDA recalls Z-1813-2009
- Recall initiated FDA recalls Z-0008-2015
- Recall initiated FDA recalls Z-0485-2015
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- CBK
- Device class
- Class II
- Regulation
- 21 CFR Part 868.5895
- Advisory committee
- Anesthesiology
- FDA decision
- ST
Overview
AI-generated summary, 19 Sep 2026
The Esprit Ventilator is a continuous-use, facility-based mechanical ventilator designed to provide respiratory support for patients. It delivers controlled or assisted ventilation through precise pressure and volume regulation, ensuring adequate oxygenation and ventilation in critical care settings.
This device is intended for use in hospital environments and is supplied as a reusable medical instrument. It adheres to MRI safety standards for facility use and is subject to FDA clearance under 510(k) regulations. The ventilator operates under strict regulatory oversight, including compliance with consensus guidelines and applicable safety requirements for life-support devices.
Similar devices
Other devices filed under FDA product code CBK (Ventilator, Continuous, Facility Use).
- ADULT VENTILATOR TESTER VT-1 by Bio-Tek Instruments, Inc.
- GE DATEX-OHMEDA ENGSTROM CARESTATION by Datex-Ohmeda Inc.
- NEWBOURNS INFANT VENTILATOR by Bourns Medical Systems, Inc.
- Patient Circuit V64-0024 by FLIGHT MEDICAL INNOVATIONS LTD
- BEAR CUB INFANT PRESSURE VENTILATOR by Allied Healthcare Products, Inc.
- SERVO-I VENTILATOR SYSTEM MODEL 64 87 800 E4073, HELIOX OPTION, MODEL 6675585 by Maquet Critical Care AB
- VOLUME VENTILATOR by Sechrist Industries Inc
- Trilogy Evo AU2110X15B by Respironics Inc.
- PORTABLE LIFECARE VENTILATOR PLV-102 by Lifecare Services, Inc.
- PORTABLE LIFECARE VENTILATOR PLV-100 by Lifecare Services, Inc.