Esprit Ventilator, Model V1000
510(k) K034040
Esprit Ventilator, Model V1000 is a medical device (FDA Class II; FDA product code CBK, Ventilator, Continuous, Facility Use) made by Respironics, Inc., cleared under FDA 510(k) K034040 on 21 Jan 2004, active as of 22 Oct 2026. 8 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK034040 | Date21 Jan 2004 | StatusActive |
Official records
- FDA 510(k) database K034040
From the FDA record
- Received by the FDA
- 29 Dec 2003
- FDA decision
- 21 Jan 2004 (Substantially Equivalent)
- FDA review time
- 23 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Terminated Z-0485-2015
If the power supply fan mounting screws are installed incorrectly, the ends of the screws may touch the power supply printed circuit board assembly (PCBA) and short to ground. This may prevent the ventilator to switch to battery power if AC power is disrupted.
-
- Terminated Z-0008-2015
A failure of a specific component may prevent the ventilators from operating on AC power or transitioning back to AC power when operating on battery. Additionally, if a battery is not present or is depleted the ventilators will not operate. This failure will cause the loss of ventilator support, which may result in hypercarbia or hypoxemia.
-
- Terminated Z-1813-2009
The recall was initiated because through routine product monitoring Respironics discovered that under certain conditions, when operating on AC line power, a portion of the current from the External Battery charger is erroneously diverted to the ventilator. When this occurs, the External Battery charger may not transition correctly. Prolonged operation in this condition results in overcharging and
-
- Terminated Z-2004-2008
- Terminated Z-2003-2008
- Terminated Z-2002-2008
This action is being taken to address power supply failures on the Esprit Ventilator which have occurred in some units. These failures, which were discovered through routine product monitoring, have been reported in countries that utilize an input voltage of greater than 200 volts AC. Some customers have reported a burning smell coming from the ventilator.
-
- Terminated Z-2084-2013
The power to the Esprit Display's Backlight is interrupted, causing the GUI to flicker or go dark and making the display unreadable.
-
- Terminated Z-1230-04
Flow sensors would exhibit drift.
Timeline
- Recall initiated FDA recalls Z-1230-04
- Cleared FDA 510(k) K034040
- Recall initiated FDA recalls Z-2084-2013
- Recall initiated FDA recalls Z-2004-2008
- Recall initiated FDA recalls Z-2003-2008
- Recall initiated FDA recalls Z-2002-2008
- Recall initiated FDA recalls Z-1813-2009
- Recall initiated FDA recalls Z-0008-2015
- Recall initiated FDA recalls Z-0485-2015
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- CBK
- Device class
- Class II
- Regulation
- 21 CFR Part 868.5895
- Advisory committee
- Anesthesiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code CBK (Ventilator, Continuous, Facility Use).
- Evolution 3E Ventilator by eVent Medical, Ltd.
- CAM-10 Portable Volume Ventilator by Lifecare Services, Inc.
- Ventoux VC2; Ventoux VC3 by FLIGHT MEDICAL INNOVATIONS LTD
- Bear 7 Ventilator by Bear Medical Systems, Inc.
- Uni-Vent Impact Minute Volume Ventilator by Impact Instrumentation, Inc.
- Modified 225/SIMV Volume Ventilator by MONAGHAN MEDICAL CORP.
- Jet Ventilator by Healthdyne, Inc.
- Servo-air 4.0 Ventilator System by Maquet Critica Care AB
- Divan Anesthesia Ventilator by Draeger Medical, Inc.
- VENTILATOR #1 by Airco/Ohio Medical Products