# Esprit Ventilator, Model V1000

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/esprit-ventilator-model-v1000-cbk>
- Manufacturer: [Respironics, Inc.](https://devicerts.com/manufacturers/respironics-inc) (United States)
- Name in the registry: ESPRIT VENTILATOR, MODEL V1000
- Device class: Class II (US FDA)
- FDA product code: CBK

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K034040 · decision 2004-01-21 · decision SESE · status: Active

## Recalls

- Z-0485-2015 · Class II · terminated · initiated 2014-10-29: If the power supply fan mounting screws are installed incorrectly, the ends of the screws may touch the power supply printed circuit board assembly (PCBA) and short to ground. This may prevent the ventilator to switch to battery power if AC power is disrupted.
- Z-0008-2015 · Class I · terminated · initiated 2014-09-17: A failure of a specific component may prevent the ventilators from operating on AC power or transitioning back to AC power when operating on battery. Additionally, if a battery is not present or is depleted the ventilators will not operate. This failure will cause the loss of ventilator support, which may result in hypercarbia or hypoxemia.
- Z-1813-2009 ·  · terminated · initiated 2009-04-22: The recall was initiated because through routine product monitoring Respironics discovered that under certain conditions, when operating on AC line power, a portion of the current from the External Battery charger is erroneously diverted to the ventilator. When this occurs, the External Battery charger may not transition correctly. Prolonged operation in this condition results in overcharging and
- Z-2004-2008 ·  · terminated · initiated 2008-01-25: This action is being taken to address power supply failures on the Esprit Ventilator which have occurred in some units. These failures, which were discovered through routine product monitoring, have been reported in countries that utilize an input voltage of greater than 200 volts AC. Some customers have reported a burning smell coming from the ventilator.
- Z-2003-2008 ·  · terminated · initiated 2008-01-25: This action is being taken to address power supply failures on the Esprit Ventilator which have occurred in some units. These failures, which were discovered through routine product monitoring, have been reported in countries that utilize an input voltage of greater than 200 volts AC. Some customers have reported a burning smell coming from the ventilator.
- Z-2002-2008 ·  · terminated · initiated 2008-01-25: This action is being taken to address power supply failures on the Esprit Ventilator which have occurred in some units. These failures, which were discovered through routine product monitoring, have been reported in countries that utilize an input voltage of greater than 200 volts AC. Some customers have reported a burning smell coming from the ventilator.
- Z-2084-2013 · Class II · terminated · initiated 2007-10-29: The power to the Esprit Display's Backlight is interrupted, causing the GUI to flicker or go dark and making the display unreadable.
- Z-1230-04 ·  · terminated · initiated 2002-02-20: Flow sensors would exhibit drift.

## Identifiers

- 510(k) Number: K034040 [FDA 510(k)]
- FDA product code: CBK [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 868.5895 [FDA 510(k)]
- Advisory committee: Anesthesiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Esprit Ventilator, Model V1000, Respironics, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/esprit-ventilator-model-v1000-cbk (data as of 2026-11-21).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-21
- JSON: <https://devicerts.com/devices/esprit-ventilator-model-v1000-cbk.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/esprit-ventilator-model-v1000-cbk
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.