Evms (Enterprise Visual Medical System Software)
510(k) K053281
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK053281 | Date7 Dec 2005 | StatusActive |
Recalls
-
- Terminated Z-0614-2017
It was reported by a customer that RadSuite images are not appearing as readable images. This product malfunction has the potential to result in delay in care and a possible re-exposure of the patient to radiation (x-ray).
-
- Terminated Z-1470-2017
The software produced a number of "do not route" exceptions, which may result in potential patient injury or delay in diagnosis or treatment.
-
- Terminated Z-2627-2016
Potential for RadSuite AV viewer to skip image slices when the user presses page up or page down when the viewer displays multiple viewports and the user has mouse focused on any viewport other than the first viewport.
-
- Terminated Z-2685-2016
Interventional Radiology (IR) images are stored as JPEG2k Lossless in Merge Enterprise Archive (EA) and are not displaying correctly in RadSuite.
-
- Terminated Z-0794-2016
Potential incorrect Standardized Uptake Values (SUV) measurements in RadSuite.
-
- Terminated Z-1180-2017
The values provided from the Pixel Value tool do not appear to be correct, which may result in potential patient injury or delay in diagnosis or treatment.
-
- Terminated Z-2715-2016
When RadSuite is used with IPID (Issuer of Patient ID) as a part of the "Patient Identifier," it is possible in some circumstances that the demographics of one patient will be applied to a study or studies for another patient.
-
- Terminated Z-2796-2016
Multiple studies were missing at a customer location that never made it to the archive before watermarking
-
- Terminated Z-1644-2008
Miscalculation: The calculation of the standard uptake value does not include the required calibration factor/decay correction factor. The resulting value is incorrectly calculated, which may lead to incorrect interpretation of tissue density. The calculation error exists in both the software and user guide.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LLZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.2050
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent