# Evotech Endoscope Cleaner and Reprocessor

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/evotech-endoscope-cleaner-and-reprocessor-feb>
- Manufacturer: [Advanced Sterilization Products](https://devicerts.com/manufacturers/advanced-sterilization-products-inc) (United States)
- Name in the registry: EVOTECH ENDOSCOPE CLEANER AND REPROCESSOR
- Device class: Class II (US FDA)
- FDA product code: FEB

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K082392 · decision 2008-11-13 · decision SESE · status: Active

## Recalls

- Z-0965-2017 · Class II · terminated · initiated 2016-09-27: Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor System because the hose may come loose and cause a leak.
- Z-1150-2013 · Class II · terminated · initiated 2013-03-28: Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because they have received reports of smoke emanating from the system due to a circuit board malfunction that may occur in the Channel Control Manifold sub-assembly.
- Z-1123-2013 · Class II · terminated · initiated 2013-03-22: Advanced Sterilization Products (ASP) has determined that a limited number of EVOTECH ECRs manufactured between 2011 and 2012 did not receive sufficient High Potential (HIPOT) DC voltage testing.

## Identifiers

- 510(k) Number: K082392 [FDA 510(k)]
- FDA product code: FEB [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 876.1500 [FDA 510(k)]
- Advisory committee: Gastroenterology, Urology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Evotech Endoscope Cleaner and Reprocessor, Advanced Sterilization Products. deviCERTS medical device catalogue. https://devicerts.com/devices/evotech-endoscope-cleaner-and-reprocessor-feb (data as of 2026-11-28).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-28
- JSON: <https://devicerts.com/devices/evotech-endoscope-cleaner-and-reprocessor-feb.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/evotech-endoscope-cleaner-and-reprocessor-feb
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.