# ExacTrac Vero 3.5

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/exactrac-vero-3-5>
- Manufacturer: [Brainlab SE](https://devicerts.com/manufacturers/brainlab-se) (Germany)
- Device class: Class II (US FDA)
- FDA product code: IYE
- GMDN: Radiological whole-body positioner (40902)

## Registrations and authorizations

- **GUDID** · GUDID listing · United States · UDI-DI 04056481132446 · listed 2016-12-19 · status: Listed (GUDID: not in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K122451 · decision 2012-09-28 · decision SESE · status: Active · from GUDID

## Recalls

- Z-0172-2018 · Class II · terminated · initiated 2017-11-03: Brainlab has internally detected that under specific conditions ExacTrac Vero may not correctly account for this ring angle correction during the calculation of the corresponding couch shift. This may result in the treatment couch not moving to the exact planned position resulting in a deviation between the planned and the treatment target position. This deviation may not be clearly visible to the user if no verification image of the patient position is acquired after this couch shift has been performed.
- Z-1929-2016 · Class II · terminated · initiated 2016-05-04: Potentially incorrect positioning when using Implanted Marker Detection with Brainlab ExacTrac Vero 3.5

## Identity

- Trade name: ExacTrac Vero [FDA GUDID (AccessGUDID)]
- Generic name: Radiological whole-body positioner [FDA GUDID (AccessGUDID)]
- Model / reference: 3.5 [FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 04056481132446 [FDA GUDID (AccessGUDID)]
- 510(k) Number: K122451 [FDA GUDID (AccessGUDID)]
- FDA product code: IYE [FDA GUDID (AccessGUDID)]
- GMDN term: Radiological whole-body positioner (40902) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 892.5050 [FDA GUDID (AccessGUDID)]
- GUDID distribution status: Not in Commercial Distribution [FDA GUDID (AccessGUDID)]

## Properties

- Lot / batch number: Yes [FDA GUDID (AccessGUDID)]
- Serial number: Yes [FDA GUDID (AccessGUDID)]
- Supplied sterile: No [FDA GUDID (AccessGUDID)]
- Sterilise before use: No [FDA GUDID (AccessGUDID)]
- Single use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [FDA GUDID (AccessGUDID)]
- Contains latex: No [FDA GUDID (AccessGUDID)]
- Prescription use: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Expiration date: No [FDA GUDID (AccessGUDID)]
- Manufacturing date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: ExacTrac Vero 3.5, Brainlab SE. deviCERTS medical device catalogue. https://devicerts.com/devices/exactrac-vero-3-5 (data as of 2026-11-22).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-23
- Statuses computed as of: 2026-11-22
- JSON: <https://devicerts.com/devices/exactrac-vero-3-5.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/exactrac-vero-3-5
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.