Exactrac Vero
by Brainlab AG
510(k) K122451
Exactrac Vero is a medical device (FDA Class II; FDA product code IYE, Accelerator, Linear, Medical) made by Brainlab AG, cleared under FDA 510(k) K122451 on 28 Sep 2012, active as of 22 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK122451 | Date28 Sep 2012 | StatusActive |
Official records
- FDA 510(k) database K122451
From the FDA record
- Received by the FDA
- 13 Aug 2012
- FDA decision
- 28 Sep 2012 (Substantially Equivalent)
- FDA review time
- 46 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Terminated Z-0172-2018
Brainlab has internally detected that under specific conditions ExacTrac Vero may not correctly account for this ring angle correction during the calculation of the corresponding couch shift. This may result in the treatment couch not moving to the exact planned position resulting in a deviation between the planned and the treatment target position. This deviation may not be clearly visible to the user if no verification image of the patient position is acquired after this couch shift has been performed.
-
- Terminated Z-1929-2016
Potentially incorrect positioning when using Implanted Marker Detection with Brainlab ExacTrac Vero 3.5
Timeline
- Cleared FDA 510(k) K122451
- Recall initiated FDA recalls Z-1929-2016
- Recall initiated FDA recalls Z-0172-2018
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- IYE
- Device class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K122451
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| ExacTrac Vero 3.5 | 3.5 | 04056481132446 | — | No longer on the market |
| ExacTrac Vero 3.6 | 20896 / 3.6 | 04056481000769 | — | No longer on the market |
Similar devices
Other devices filed under FDA product code IYE (Accelerator, Linear, Medical).
- GILL-THOMAS-COSMAN(GTC) DENTAL TRAY by Radionics Software Applications, Inc.
- MOSAIQ Oncology Information System by ELEKTA SOLUTIONS AB
- Raycast High Precision Lung Board Solution by Orfit Industries NV
- Klarity Medical R550-HS1 by Klarity Medical Products, LLC
- VARIAN CLINAC 1800S by Varian Assoc., Inc.
- IMAGERT by Impac Medical Systems, Inc.
- STANDARD BREAST BOARDS; DELUXE BREAST BOARD by Bionix Development Corp.
- ArcherQA (V1.0) by Wisdom Technologies., Inc.
- KLARITY MEDICAL RNE-103PL by Klarity Medical Products, LLC
- BAT ULTRASOUND LOCALIZATION AND POSITIONING SYSTEM, VERSION 6 by Nomos Corp.