Fabius Gs/Fabius Tiro Anesthesia System

by Draeger Medical, Inc.

510(k) K042419

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

11 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK042419 Date8 Oct 2004ClearedStatusActive

Recalls

11 recalls · FDA

  • 2 recalls, same reason

    • Terminated Z-2250-2018 FDA Class II initiated 24 May 2018
    • Terminated Z-2251-2018 FDA Class II initiated 24 May 2018

    The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.

  • 3 recalls, same reason

    • Terminated Z-1484-2018 FDA Class I initiated 1 Mar 2018
    • Terminated Z-1482-2018 FDA Class I initiated 1 Mar 2018
    • Terminated Z-1483-2018 FDA Class I initiated 1 Mar 2018

    The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.

    • Terminated Z-0008-2014 FDA Class I initiated 14 Aug 2013

    During final testing of the Draeger Fabius GS Premium, Fabius OS, and Fabius Tiro anesthesia machines, some units did not pass the high voltage test and the minimum clearance between an electrical component and the unit housing was not maintained for some from a specific batch.

  • 2 recalls, same reason

    • Terminated Z-2478-2010 FDA Unclassified (pre-2012) initiated 26 Aug 2010
    • Terminated Z-2479-2010 FDA Unclassified (pre-2012) initiated 26 Aug 2010

    potential failures to Electri-Cord Manufacturing AC power cord

    • Terminated Z-2338-2008 FDA Unclassified (pre-2012) initiated 7 Jul 2008

    The parent company became aware of instances where the lower side rail of the frame of the machine broke inwards when being moved across a threshold; resulting in the machine being unstable.

    • Terminated Z-1315-2008 FDA Unclassified (pre-2012) initiated 5 Feb 2008

    Detachment of Component: The chassis on some machines were manufactured out of specifications. As a result, the caster may break loose from the chassis.

    • Terminated Z-0048-2007 FDA Unclassified (pre-2012) initiated 3 Oct 2006

    One of four casters may break loose from chassis

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
BSZ
Device Class
Class II
Regulation
21 CFR Part 868.5160
Advisory Committee
Anesthesiology
FDA Decision
SESE Substantially Equivalent

Other devices from Draeger Medical, Inc.

See all devices from Draeger Medical, Inc. →