Fabius MRI
by Draeger Medical AG & Co. KG 510(k)
Active
510(k) K072884
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK072884 | Date30 Jul 2008 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-0501-2022
26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas flow could deviate from the adjusted value. If the deviation is not noticed, insufficient oxygenation of the patient may occur.
-
- Terminated Z-2252-2018
The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.
-
- Terminated Z-1481-2018
The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.
-
- Terminated Z-2271-2015
the Fabius MRI or parts of the system were attracted by the magnetic field of the MRI. If the Fabius MRI is positioned too close to the MRI, its parts can be loosened as a result of the magnetic force of the MRI.
Specification
- 510(k) Number
- FDA Product Code
- BSZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 868.5160
- Advisory Committee
- Anesthesiology
- FDA Decision
- SESE Substantially Equivalent