m2000, product family

by Abbott Molecular 7 variants Basic UDI-DI 0884999RTN0106AC

Shared by every variant

Manufacturer
Abbott Molecular(7)
Basic UDI-DI
0884999RTN0106AC
Risk class
Class IClass A (varies)
Legislation
IVDR
Market status
7 on the market
Official record
EUDAMED public site — Open the EUDAMED public site, choose Devices, and search for the Basic UDI-DI 0884999RTN0106AC.

Recalls

7 recalls · FDA

  • Z-0463-2015 Class II · terminated · 23 Oct 2014
    Some versions of Application Specifications (App Spec) are incompatible with m2000sp system software version 6.0 and 7.0. This issue will cause Error Code 9000 (An unexpected error has occurred: Software error) to be generated when the Available Reagent Packs screen is selected.
  • Z-2169-2014 Class II · terminated · 19 Jun 2014
    Abbott Molecular identified that some of the 5 ml Reaction Vessels (RVs) contained in lots 56944001, 57026001, 57059001, and 57673001 (original vendor lot 2198077) may have an insufficient rim. An RV with an insufficient rim may not be held securely in the 1 mL Subsystem Carrier. This may cause the RV to be seated incorrectly in the carrier or drop through it.
  • Z-0948-2015 Class II · terminated · 23 Nov 2013
    The x-drive assembly may fail prematurely. This could cause the manipulator or liquid handling arm to contact an object, shake and drip, or improperly pipet.
  • Z-0634-2013 Class II · terminated · 4 Dec 2012
    An issue with the m2000sp Instrument software has been identified whereby, after a series of steps, if changes are made to the control or calibrator information on the Sample Extraction Assay Details screen, the changes are not saved.
  • Z-2013-2012 Class II · terminated · 31 Jan 2012
    In the Abbott Molecular Bar Code Scanner User's Guide, List Number 6L88-02/09, the section Modifying the Keyboard Format for Bar Code Scanner LN 6L89-01 contains incorrect bar codes that can prevent the correct configuration of the bar code scanner language format. The Abbott m2000 System is intended for use in performing nucleic acid testing in clinical laboratories. It is comprised of the Ab
  • Z-0365-2012 unclassified · terminated · 11 Oct 2011
    Abbott Molecular has received reports in which the liquid waste sensor and components in the m2000sp lower instrument cabinet overheated and/or caught fire.
  • Z-0048-2012 unclassified · terminated · 8 Sep 2011
    The Abbott m2000rt System software has software bug which can affect customers who manually enter a Test Order onto the m2000rt and sorts a column on the calibration or control grids of the Test Order screen by double clicking the column header. This causes a mismatch between specific concentration values and the designated calibrators and controls in the final results, which could cause an inval

Recalls matched to the 7 devices listed on this page; other pages of the family can carry others.

Variants

7 devices · page 1 of 1

Variants of m2000, Basic UDI-DI 0884999RTN0106AC
UDI-DISize and differences
m2000 01N77-0100884999000995 —On the market
m2000 05N92-00100884999015463 —On the market
m2000 01N06-01 01N06-01 / 01N06-0100884999000414 —On the market
m2000 01N97-01 01N97-01 / 01N97-0100884999001381 —On the market
m2000 04J72-82 04J72-82 / 04J72-8200884999006898 —On the market
m2000 04J72-86 04J72-86 / 04J72-8600884999006911 —On the market
m2000 9K31-01 9K31-01 / 9K31-0100884999032057 —On the market