Farastar™
by Boston Scientific Corporation 27 variants
Ref M004PF61M412
Active
CE MDR
Valid until
Not listed
FDA recalls
5 ongoingDocuments
None linked in EUDAMED
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number01915060000000000000537NG 00191506042493 | Date22 May 2025 | StatusActive |
| United States | GUDID listing GUDID | Number00191506042493 | Date9 Feb 2024 | StatusListed |
Certificates
EUDAMED lists no certificate for this device's Basic UDI-DI.
Recalls
-
- Ongoing Z-2753-2026
Boston Scientific is replacing specific FARASTAR PFA generators which have a configuration that is not compatible for use with the FARAPOINT PFA catheter.
-
- Ongoing Z-2750-2026
- Ongoing Z-2752-2026
- Ongoing Z-2751-2026
Labeling discrepancy: Devices distributed with Instructions For Use (IFU) did not include FARAPOINT PFA catheter information.
-
- Ongoing Z-2278-2025
The potentially impacted units were manufactured using specific equipment that may have caused cracks in the electrode bands on the catheter's distal end.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Generic Name
- Electrical-only medical device connection cable, reusable
- Device Name
- FARAPULSE PFA Capital Equipment Components
- Model / Ref / Catalog
- M004PF61M412
- Primary DI / UDI-DI
- 00191506042493
- Basic UDI-DI
- 01915060000000000000537NG
- PMA Number
- FDA Product Code
- QZI
- EMDN Code
- C020399 Cardiac tissue ablation devices - other
- GMDN Term
- Electrical-only medical device connection cable, reusable
- Device Class
- Class I EUDAMEDClass III GUDID
- EU market of registration
- Netherlands
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device Class | Class I | Class III |
Manufacturer details
- EUDAMED SRN
- US-MF-000004702
- Authorised representative
- Guidant Europe NV/SA; Boston Scientific BE-AR-000008410
Other devices from Boston Scientific
Properties
Other properties (20)
- Sterile: No
- Single Use: No
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Measuring Function: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Expiration Date: No