Skip to content

Fdr Aqro (Dr-Xd 1000)

by Fujifilm Medical Systems U.S.A, Inc.

510(k) K170858

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK170858 Date24 Apr 2017ClearedStatusActive

Recalls

2 recalls · 2 ongoing · FDA

    • Ongoing Z-3158-2024 FDA Class II initiated 23 Nov 2023

    When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.

    • Ongoing Z-1470-2023 FDA Class II initiated 27 Feb 2023

    Arm which holds the tube head in place is susceptible to breaking, if the device breaks it could potentially cause serious harm or injury to the patient and/or user.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
IZL
Device Class
Class II
Regulation
21 CFR Part 892.1720
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent