Skip to content

Fibrinogen Screen

by Pacific Hemostasis

510(k) K800826

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

2 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 11 Oct 2026. What this means

Regulatory status

as of registry sync 11 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK800826 Date30 May 1980ClearedStatusActive

Recalls

2 recalls · FDA

  • 2 recalls, same reason

    • Terminated Z-2762-2017 FDA Class II initiated 19 May 2017
    • Terminated Z-2763-2017 FDA Class II initiated 19 May 2017

    An issue was discovered with the Pacific Hemostasis Fibrinogen Degradation Products (FDP) kit, list number 100650, lot 948546 and FDP Latex Reagent, list 100651, lot 890199. While the kit performs accurately at FDP concentrations greater than 20 g/ml, there is an observed reduction in sensitivity when FDP levels are less than 20 g/ml. This reduction in sensitivity has been tied to FDP Latex Reagent lot 890199 and may result in false negatives.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
KQJ
Device Class
Class II
Regulation
21 CFR Part 864.7340
Advisory Committee
Hematology
FDA Decision
SESE Substantially Equivalent

Other devices from Pacific Hemostasis