Filtered Extension Sets

by B. Braun of America, Inc.

510(k) K942988

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

19 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK942988 Date14 Jul 1995ClearedStatusActive

Recalls

32 recalls · 19 ongoing · FDA

    • Ongoing Z-0193-2027 FDA Not yet classified initiated 14 Sep 2026

    Potential for leakage and backflow at the Safsite valve.

    • Ongoing Z-3089-2026 FDA Class II initiated 4 Aug 2026

    Specific lots of Discofix Stopcock Sets recalled due to an incorrect expiration date printed on the carton shipper label.

  • 15 recalls, same reason

    • Ongoing Z-0653-2026 FDA Class II initiated 29 Oct 2025
    • Ongoing Z-0692-2026 FDA Class II initiated 29 Oct 2025
    • Ongoing Z-0604-2026 FDA Class II initiated 29 Oct 2025
    • Ongoing Z-0663-2026 FDA Class II initiated 29 Oct 2025
    • Ongoing Z-0628-2026 FDA Class II initiated 29 Oct 2025
    • Ongoing Z-0704-2026 FDA Class II initiated 29 Oct 2025
    • Ongoing Z-0634-2026 FDA Class II initiated 29 Oct 2025
    • Ongoing Z-0633-2026 FDA Class II initiated 29 Oct 2025
    • Ongoing Z-0626-2026 FDA Class II initiated 29 Oct 2025
    • Ongoing Z-0703-2026 FDA Class II initiated 29 Oct 2025
    • Ongoing Z-0687-2026 FDA Class II initiated 29 Oct 2025
    • Ongoing Z-0716-2026 FDA Class II initiated 29 Oct 2025
    • Ongoing Z-0632-2026 FDA Class II initiated 29 Oct 2025
    • Ongoing Z-0715-2026 FDA Class II initiated 29 Oct 2025
    • Ongoing Z-0635-2026 FDA Class II initiated 29 Oct 2025

    Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

    • Ongoing Z-0092-2023 FDA Class II initiated 15 Sep 2022

    The extension set has a high probability of potential failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may expose both patients and healthcare workers to hazardous medications. Depending on the nature of the leak and type of infusion, there is also the potential for life-threatening situations to occur including air embolization or hypovolemia.

    • Ongoing Z-0091-2023 FDA Class II initiated 15 Sep 2022

    The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may expose both patients and healthcare workers to hazardous medications. Depending on the nature of the leak and type of infusion, there is also the potential for life-threatening situations to occur including air embolization or hypovolemia.

    • Terminated Z-0727-2022 FDA Class II initiated 31 Jan 2022 on shared 510(k) K942988

    This lot may have an elevated risk of leakage from the Air-Inlet Filter of the device. Leakage of the Air-Inlet Filter may increase the risk of decreasing the dose of medication delivered to the patient, cause delays in treatment due to the need to replace the device or may cause contamination of the fluid path that can lead to bloodstream infection.

  • 2 recalls, same reason

    • Terminated Z-2552-2021 FDA Class II initiated 10 Aug 2021
    • Terminated Z-2546-2021 FDA Class II initiated 10 Aug 2021

    Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing and bloodstream infections.

    • Terminated Z-1509-2021 FDA Class II initiated 15 Mar 2021

    BBMI has identified through complaints the potential for the extension set to become occluded or unable to prime.

    • Terminated Z-2022-2014 FDA Class II initiated 29 May 2014

    Complaints were received that reported the stopcock of some Stopcock Extension Sets are assembled in the reverse position, resulting in the stopcock being closed when set in the open position and vise versa.

    • Terminated Z-0987-2014 FDA Class II initiated 8 Jan 2014

    Complaints were received reporting leakage on certain lots of SAFSITE Injection Sites and IV Sets manufactured with SAFSITE Injection Sites.The SAFSITE Injection Site is intended to provide needle free access on IV and Extension Sets.

  • 5 recalls, same reason

    • Terminated Z-0673-2010 FDA Unclassified (pre-2012) initiated 24 Nov 2009
    • Terminated Z-0670-2010 FDA Unclassified (pre-2012) initiated 24 Nov 2009
    • Terminated Z-0672-2010 FDA Unclassified (pre-2012) initiated 24 Nov 2009
    • Terminated Z-0671-2010 FDA Unclassified (pre-2012) initiated 24 Nov 2009
    • Terminated Z-0674-2010 FDA Unclassified (pre-2012) initiated 24 Nov 2009

    System error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the set to be used. This requires a new set to be obtained, thus creating a potential for delay in therapy.

    • Terminated Z-2408-2008 FDA Unclassified (pre-2012) initiated 30 Jun 2008 on shared 510(k) K942988

    incorrect burette was packaged with the product

    • Terminated Z-0933-03 FDA Unclassified (pre-2012) initiated 14 Apr 2003

    possible leakage

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
FPA
Device Class
Class II
Regulation
21 CFR Part 880.5440
Advisory Committee
General Hospital
FDA Decision
SESE Substantially Equivalent