Skip to content

FlowGuide FG60F

by IMDS Operations B.V.

FlowGuide FG60F is a vascular guide-catheter (FDA Class II; FDA product code DQY, Catheter, Percutaneous) made by IMDS Operations B.V., cleared under FDA 510(k) K242337 on 17 Apr 2025 and listed in FDA GUDID since 1 Jun 2026 (UDI-DI 08718481740347), active as of 23 Oct 2026.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None found

in data synced 23 Oct 2026

Documents

510(k) K242337

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number08718481740347 UDI-DI Date1 Jun 2026ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK242337 Date17 Apr 2025DecidedStatusActive

Official records

From the FDA record

Device description
The FlowGuide is Rapid Exchange or so-called Monorail design. The FlowGuide has a distal shaft connected to a proximal shaft / push rod with a proximal end stop connected at the proximal end. The distal shaft is stainless-steel reinforced. The reinforced distal shaft ends in an integrated soft distal tip to facilitate advancement of the device through a guiding catheter and into the main coronary vessels. In addition, the outer surface of the distal shaft and tip are equipped with a hydrophilic coating and the inner surface with a silicon coating for increased lubricity. The distal shaft has a radiopaque distal tip that in conjunction with fluoroscopy, aids in the placement of the guide extension. The distal shaft of the FlowGuide has an additional radiopaque marker at the proximal end that in conjunction with fluoroscopy, aid in preventing distal shaft exit the guiding catheter. Furthermore, the FlowGuide has side holes in its distal shaft to allow blood flow through the side holes to distal vasculature to aid in reducing the risk of ischemia. The guide extension has two markers on the push rod that indicate approximately, the exit of the distal tip from the guiding catheter (brachial – 90 cm, femoral – 100 cm). The IMDS family of guide extensions will be used in guiding catheters of 90 and 100 cm length, supplied with accessories. The effective length of the guide extensions will be 150 cm. For FlowGuide, a flexible transition zone is present on the push rod as an additional aid for smooth entry of intravascular devices.
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied sterile

From the GUDID record (AccessGUDID), as read on 23 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA 510(k) K242337 decided 17 Apr 2025 Active
  • FDA GUDID UDI-DI 08718481740347 listed 1 Jun 2026 Listed

Timeline

2 dated events

  1. Cleared FDA 510(k) K242337
  2. Listed FDA GUDID 08718481740347

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
FlowGuide
Generic name
Vascular guide-catheter, single-use
Model / reference
FG60F
Primary DI / UDI-DI
08718481740347
510(k) Number
FDA product code
DQY
GMDN term
Vascular guide-catheter, single-use
Device class
Class II
Regulation number
870.1250
GUDID distribution status
In Commercial Distribution

Similar devices

Other devices filed under FDA product code DQY (Catheter, Percutaneous).

Other devices from IMDS Operations B.V.

  • Guidion Short GS60F

    Class II
  • TrapIT TRP9015

    Class II
  • ShapeIT SI014135

    Class II
  • TrapIT TRP10015

    Class II

See all devices from IMDS Operations B.V. →

Properties

  • Supplied sterile
  • Single use
  • Prescription use
  • Lot / batch number
  • Expiration date
  • Manufacturing date
Other properties (9, all answered No)