Fort Defiance Automated Steam Sterilizer
by Fort Defiance Industries, Inc. 510(k)
Active Open recall
510(k) K141009
Regulatory status
Recall ongoing. 2 ongoing FDA recalls name this device. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK141009 | Date16 Dec 2014 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-2983-2026
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
-
- Ongoing Z-0121-2022
The rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground.
-
- Terminated Z-2411-2018
During long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the plumbing of the P2131 device produces deposits in the water system that can potentially interfere with the rear sight gauge and / or the water liquid level switch.
Specification
- 510(k) Number
- FDA Product Code
- FLE
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.6880
- Advisory Committee
- General Hospital
- FDA Decision
- SESE Substantially Equivalent