Fresenius 2008T Hemodialysis Machine with Bibag System
by Fresenius Medical Care North America
510(k) K121341
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK121341 | Date6 Dec 2012 | StatusActive |
Recalls
-
- Terminated Z-0829-2018
While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.
-
- Terminated Z-1325-2016
Added Warning Statement: Fresenius 2008 Series Hemodialysis Machines False Blood Leak Alarm When Dialyzing Patients Treated with Hydroxocobalamin (or any form of Vitamin B-12)
-
- Terminated Z-1550-2015
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
-
- Terminated Z-1015-2015
Unqualified power assemblies shipped to customers.
-
- Terminated Z-1618-2014
2008-Series Hemodialysis Machines with Revision K Actuator Test Boards may fail to detect a disconnected Ultrafiltration (UF) Pump
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- KDI
- Device Class
- Class II
- Regulation
- 21 CFR Part 876.5860
- Advisory Committee
- Gastroenterology, Urology
- FDA Decision
- SESE Substantially Equivalent