  "format": "devicerts.device.v1",
  "cite_as": "Galaxy System, Noah Medical Corp. deviCERTS medical device catalogue. https://devicerts.com/devices/galaxy-system-eoq (data as of 2026-11-22).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/galaxy-system-eoq",
    "json_url": "https://devicerts.com/devices/galaxy-system-eoq.json",
    "markdown_url": "https://devicerts.com/devices/galaxy-system-eoq.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-22",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/galaxy-system-eoq",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Galaxy System",
    "registry_name": "Galaxy System",
    "url": "https://devicerts.com/devices/galaxy-system-eoq",
    "manufacturer": {
      "name": "Noah Medical Corp.",
      "url": "https://devicerts.com/manufacturers/noah-medical-corp",
      "country": null,
      "country_basis": null,
      "fda_registration_country": "United States"
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K223144",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "EOQ"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K223144",
      "number_label": null,
      "device_udi_di": null,
      "date": "2023-03-01",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0598-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-09-12",
      "reason": "Due to Users not having access to the Instructions for Use/User Manual due to it being password protected.",
      "applies": "this device",
      "via": "K223144"
    },
    {
      "number": "Z-0742-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-10-28",
      "reason": "Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increase the chance of injury to the tissues of the lung as the scope is withdrawn including the possibility of pneumothorax.",
      "applies": "this device",
      "via": "K223144"
    },
    {
      "number": "Z-0693-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-10-25",
      "reason": "Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.",
      "applies": "this device",
      "via": "K223144"
    },
    {
      "number": "Z-0679-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-10-25",
      "reason": "Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.",
      "applies": "this device",
      "via": "K223144"
    },
    {
      "number": "Z-1734-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-03-09",
      "reason": "Due to a misaligned calibration of the wired controller that may result in unintended motion of the scope.",
      "applies": "this device",
      "via": "K223144"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K223144",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "EOQ",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 874.4680",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Ear, Nose, Throat",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}