Gammamed 12I
by Mick Radio-Nuclear Instruments, Inc.
510(k) K891131
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK891131 | Date7 Aug 1989 | StatusActive |
Recalls
-
- Terminated Z-0330-2012
- Terminated Z-0329-2012
The Product Models 1 & 2 have a design flaw problem where the clamping adapter can obstruct movement of the source cable in the event the bronchial catheter is dislodged from the connector.
-
- Terminated Z-0486-2007
Medical device for patient treatment does not meet electrical safety standards.
-
- Terminated Z-1398-06
Software control program for a medical device used in radiation treatment may cause practitioners to incorrectly administer the treatment plan to cancer patients. There is a hazard when entering the treatment plans manually that the user neglects to change the default step size and/or the origin or accidentally enters incorrect parameters.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- JAQ
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.5700
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent