# GE Precision Mpi

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/ge-precision-mpi-owb>
- Manufacturer: [Nrt - Nordisk Røntgen Teknik A/S](https://devicerts.com/manufacturers/nrt-nordisk-rontgen-teknik-a-s) (Denmark)
- Name in the registry: GE PRECISION MPI
- Device class: Class II (US FDA)
- FDA product code: OWB

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K033486 · decision 2003-11-19 · decision SESE · status: Active

## Recalls

- Z-2123-2015 · Class II · terminated · initiated 2015-06-12 · matched through shared submission K033486, may concern other models: A reported incident of a monitor that fell from the suspension. A fall of a monitor could result in an injury to a patient or operator. Recall affects the single and dual LCD monitor suspensions used on Precision 500D Classical R & F system, Precision RXi 23A/32A Analog System, Precision RXi Digital system and Precision MPi fluoroscopic imaging systems.
- Z-0911-2014 · Class II · terminated · initiated 2013-12-12: GE Healthcare has recently become aware of a potential safety issue involving the Precision 500D, Precision MPI & Precision RXI X-ray imaging systems overhead LCD Video Monitor Suspension (Single monitor & Dual monitor). There is a potential for the locking nut to loosen out of the LCD Video Monitor Suspension yoke. This will not lead to an immediate fall of the suspended monitor since it will
- Z-1658-2009 ·  · terminated · initiated 2009-05-01 · matched through shared submission K033486, may concern other models: The GE Precision MPi is in violation with 21 CFR Part 1020.32 requiring a signal audible to the fluoroscopist to sound for each passage of 5 minutes of fluoroscopic irradiation time during an examination or procedure.
- Z-0753-2016 · Class II · terminated · initiated 2008-07-01: It was discovered that the Remote Touch Panel (RTP) of the GE Precision MPi X-ray system may not always boot up as intended and needs to be updated to properly accomplish its intended purpose.
- Z-2143-2008 ·  · terminated · initiated 2008-05-01 · matched through shared submission K033486, may concern other models: The Precision MPi collimator did not contain a label identifying the device as being certified to comply with applicable requirements of the x-ray performance standard.
- Z-2144-2008 ·  · terminated · initiated 2008-05-01 · matched through shared submission K033486, may concern other models: The Precision MPi collimator did not contain a label identifying the device as being certified to comply with applicable requirements of the x-ray performance standard.
- Z-2142-2008 ·  · terminated · initiated 2008-05-01 · matched through shared submission K033486, may concern other models: The Precision MPi collimator did not contain a label identifying the device as being certified to comply with applicable requirements of the x-ray performance standard.

## Identifiers

- 510(k) Number: K033486 [FDA 510(k)]
- FDA product code: OWB [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 892.1650 [FDA 510(k)]
- Advisory committee: Radiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: GE Precision Mpi, Nrt - Nordisk Røntgen Teknik A/S. deviCERTS medical device catalogue. https://devicerts.com/devices/ge-precision-mpi-owb (data as of 2026-11-16).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-16
- JSON: <https://devicerts.com/devices/ge-precision-mpi-owb.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/ge-precision-mpi-owb
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.