GE Signa 3.0T with Excite MR System
510(k) K040444
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK040444 | Date14 Apr 2004 | StatusActive |
Recalls
-
- Ongoing Z-0614-2022
The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
-
- Terminated Z-0112-2019
- Terminated Z-0111-2019
- Terminated Z-0110-2019
The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.
-
- Terminated Z-0256-2016
When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.
-
- Terminated Z-2339-2015
A gradient cable in the ACGD cabinet may be missing a spacer that prevents connecting the gradient cable terminals in the wrong polarity on X gradient amplifier. This could cause images to be flipped left-right with incorrect orientation annotation.
-
- Terminated Z-1305-2015
GE Healthcare has become aware of a potential safety issue involving MRI systems due to software versions not being maintained properly at some sites.
-
- Terminated Z-1035-2015
At certain sites, the MRU may not be connected to the magnet. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The MRU must be connected to the magnet at all times.
-
- Terminated Z-1943-2013
Screen save images are included in wrong exam due to duplicate UID creation. This may cause a safety issue due to a potential patient data mismatch.While the issue has only been observed at one customer site on the Signa HDxT 1.5T system, internal investigation revealed that the issue also impacts multiple GE healthcare MR Systems.
-
- Terminated Z-0815-2007
To provide the 3.0T MRI systems users with the 8-channel brain coil the proper coil weighting software. The proper coil weighting factors for the eight-channel brain coil were not included in the software.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LNH
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1000
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent