GE Voluson E8 Ultrasound System
510(k) K061682
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK061682 | Date12 Jul 2006 | StatusActive |
Recalls
-
- Terminated Z-0541-2010
GE Healthcare has recently become aware of an incorrect operating procedure associated with software versions 9.0.0, 9.0.1, and 9.0.2 of the Voluson E6 or Voluson E8 Diagnostic Ultrasound Systems that may impact patient safety. When using the combination of 2D-CRI and PW with a steered Doppler angle, the displayed velocity scale is incorrect. The consequence can be an underestimation of flow v
-
- Terminated Z-2404-2008
Inaccurate dimension measurements in M-Mode when pan-Zoom is activated in GE Voluson E8 Ultrasound Systems. Of particular concern are potential inaccuracies of fetal cardiac measurements that may impact patient safety.
-
- Terminated Z-0113-2008
When a spectrum of a Pulsed Wave Doppler shows an angle correction unequal to 0 is reloaded from the archive or the image is sent to a DICOM station, this angle correction is mistakenly disregarded resulting in an underestimation of the flow velocities.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- IYO
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1560
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent