  "format": "devicerts.device.v1",
  "cite_as": "Gemini Condor, Philips Medical Systems (Cleveland), Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/gemini-condor-kps (data as of 2026-11-20).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/gemini-condor-kps",
    "json_url": "https://devicerts.com/devices/gemini-condor-kps.json",
    "markdown_url": "https://devicerts.com/devices/gemini-condor-kps.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-20",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/gemini-condor-kps",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Gemini Condor",
    "registry_name": "GEMINI CONDOR",
    "url": "https://devicerts.com/devices/gemini-condor-kps",
    "manufacturer": {
      "name": "Philips Medical Systems (Cleveland), Inc.",
      "url": "https://devicerts.com/manufacturers/philips-medical-systems-cleveland-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K081135",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "KPS"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K081135",
      "number_label": null,
      "device_udi_di": null,
      "date": "2008-04-29",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1989-2021",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2021-05-01",
      "reason": "Significant wear of the Monitor Bracket Assembly shaft  goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the\ncables resulting in patient, operator, bystander, or service person  to serious injury.",
      "applies": "this device",
      "via": "K081135"
    },
    {
      "number": "Z-2199-2015",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-07-01",
      "reason": "Philips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue.",
      "applies": "this device",
      "via": "K081135"
    },
    {
      "number": "Z-1789-2015",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-03-06",
      "reason": "The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.",
      "applies": "this device",
      "via": "K081135"
    },
    {
      "number": "Z-2037-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2014-06-18",
      "reason": "When beginning a procedure to start the gantry and CT Host, a Close Estop message may appear to the operator, but disappears before action can be taken. If this occurs it results in an inability to close E-Stop and the system is not operational for clinical use. This problem occurs randomly.",
      "applies": "this device",
      "via": "K081135"
    },
    {
      "number": "Z-0979-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-12-05",
      "reason": "The TF Big Bore patient table experienced mechanical binding during horizontal table motion resulting in an automatic Emergency Stop (E-stop).  An E-Stop will interrupt a current scan, resulting in an incomplete study.",
      "applies": "this device",
      "via": "K081135"
    },
    {
      "number": "Z-0932-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-12-04",
      "reason": "The PET Reconstruction Server (PRS) database may randomly lock up after the low dose CT acquisition is complete and will not allow the PET acquisition to begin, resulting in an unexpected incomplete study.",
      "applies": "this device",
      "via": "K081135"
    },
    {
      "number": "Z-0883-2013",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-10-24",
      "reason": "Software anomaly.  Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Project",
      "applies": "this device",
      "via": "K081135"
    },
    {
      "number": "Z-2239-2012",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-08-08",
      "reason": "Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole through the casing on one of the batteries. Supplier of the UPS, Chloride/Emerson, determined the batteries, housed within the UPS battery cabinet, entered thermal runaway causing the battery to overheat.",
      "applies": "this device",
      "via": "K081135"
    },
    {
      "number": "Z-1369-2012",
      "classification": "",
      "status": "terminated",
      "initiated": "2012-01-26",
      "reason": "TumorLOC will save the incorrect contour when the user saves the contour while in the edit mode, and the software does not force the user to exit the edit mode before saving.  In TumorLOC, when saving while a contour is in edit mode, that contour will shift to the slice corresponding to the most recently drawn contour of the last organ in the Organ list.",
      "applies": "this device",
      "via": "K081135"
    },
    {
      "number": "Z-3201-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-05-31",
      "reason": "The firm is conducting a recall/software correction to several of their GEMINI PET/CT X-ray system units in the field.  The firm believes that if these software defects were to recur, it could impact the performance of reliability of the these PET/CT system units.",
      "applies": "this device",
      "via": "K081135"
    },
    {
      "number": "Z-0343-2012",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-25",
      "reason": "In response to complaints received by Philips Medical concerning certain identified software errors reported by their customers, the firm is conducting a recall/software correction to several of their GEMINI PET/CT X-ray system units in the field. The firm believes that if these software defects were to recur, it could impact the performance of reliability of the these PET/CT system units.",
      "applies": "this device",
      "via": "K081135"
    },
    {
      "number": "Z-0344-2012",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-25",
      "reason": "In response to complaints received by Philips Medical concerning certain identified software errors reported by their customers, the firm is conducting a recall/software correction to several of their GEMINI PET/CT X-ray system units in the field. The firm believes that if these software defects were to recur, it could impact the performance of reliability of the these PET/CT system units.",
      "applies": "this device",
      "via": "K081135"
    },
    {
      "number": "Z-2915-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-08-05",
      "reason": "Two non-safety related defects have been found on the Philips GEMINI TF Big Bore running version 3.6; if these defects were to recur, it could impact the performance or reliability of the system.  Philips is issuing a software update to all Philips GEMINI TF Big Bore customers to correct these defects.",
      "applies": "this device",
      "via": "K081135"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K081135",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "KPS",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.1200",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}