Genesis 71210001
Listed
Valid until
Not applicable
FDA recalls
1 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number03596010185723 | Date25 Aug 2015 | StatusListed |
Recalls
-
- Ongoing Z-0878-2020
Multiple lots of LEGION AP Femoral Cutting Blocks were manufactured with an oversize green plunger hole and therefore do not meet their design specifications.
-
- Terminated Z-0200-2019
One lot of the Legion AP femoral cutting block, size 6, was manufactured with an oversized spring.
-
- Terminated Z-2163-2018
One lot of GENESIS II 13 mm tibial punches are 2 mm smaller than specification.
-
- Terminated Z-1902-2018
A single lot of GENESIS II FEMORAL FIVE DEGREE VALGUS BUSHING due to a laser mark labeling error. The left valgus bushing was incorrectly laser marked as right and vice versa.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- Genesis
- Generic Name
- Orthopaedic inorganic implant inserter/extractor, reusable
- Model / Ref / Catalog
- 71210001
- Primary DI / UDI-DI
- 03596010185723
- 510(k) Number
- FDA Product Code
- HRY
- GMDN Term
- Orthopaedic inorganic implant inserter/extractor, reusable
- Device Class
- Class II
- Regulation Number
- 888.3530
- GUDID distribution status
- In Commercial Distribution
- Sterilization Method
- Moist Heat or Steam Sterilization
Other devices from Smith & Nephew, Inc.
Properties
Other properties (12)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Expiration Date: No
- Manufacturing Date: No