# Genesis Imrt Linear Accelerator

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/genesis-imrt-linear-accelerator-iye>
- Manufacturer: [Siemens Medical Solutions USA, Inc.](https://devicerts.com/manufacturers/siemens-medical-solutions-usa-inc) (United States)
- Name in the registry: GENESIS IMRT LINEAR ACCELERATOR
- Device class: Class II (US FDA)
- FDA product code: IYE

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K982502 · decision 1998-07-31 · decision SESE · status: Active

## Recalls

- Z-1059-2015 · Class II · terminated · initiated 2014-12-19: A hardware update to replace the old PCB Assembly, Head Driver-G41 with PCB with overcurrent protection as a result of reports of the electric board inside the gantry showing signs of overheating, including thermal damage and smoke emission.
- Z-2174-2014 · Class II · terminated · initiated 2014-05-28: A safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-sequenced treatment have the potential to lead to a collision with the patient.
- Z-1663-2013 · Class II · terminated · initiated 2013-05-29: The information regarding the FILM Mode calibration may have been considered difficult to interpret when translated. SIEMENS HEALTHCARE CR / RADIATION ONCOLOGY had become aware that this issue may have resulted in mistreatment of patients.
- Z-2639-2011 ·  · terminated · initiated 2010-11-04: The firm has become aware of a potential issue that may result in a service engineer receiving an electrical shock should the connector cable housing become unfastened from the chassis.

## Identifiers

- 510(k) Number: K982502 [FDA 510(k)]
- FDA product code: IYE [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 892.5050 [FDA 510(k)]
- Advisory committee: Radiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Genesis Imrt Linear Accelerator, Siemens Medical Solutions USA, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/genesis-imrt-linear-accelerator-iye (data as of 2026-11-23).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-23
- JSON: <https://devicerts.com/devices/genesis-imrt-linear-accelerator-iye.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/genesis-imrt-linear-accelerator-iye
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.