Gentamicin Flex Reagent Cartridge
by Dade Intl., Inc. 510(k)
Active
510(k) K962819
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK962819 | Date13 Sep 1996 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-2182-2018
The affected lots may exhibit inaccuracy for Quality Control and patient samples at the low end of the Analytical Measurement Range. The bias is due to the calibration curve not showing separation between the Level 1 and Level 2 calibrators. The negative bias observed for patient, QC, and calibrator samples, when using any of the listed lots, has the potential to impact the interpretation of gentamicin results during therapeutic monitoring at concentrations up to approximately 2.5 g/mL [5.4 mol/L].
-
- Terminated Z-2181-2018
The affected lots may exhibit inaccuracy for Quality Control and patient samples at the low end of the Analytical Measurement Range. The bias is due to the calibration curve not showing separation between the Level 1 and Level 2 calibrators.
Specification
- 510(k) Number
- FDA Product Code
- LCD
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.3450
- Advisory Committee
- Clinical Toxicology
- FDA Decision
- SESE Substantially Equivalent