Skip to content

Giraffe OmniBed Carestation CS1

by Ohmeda Medical, A Division of Datex-Ohmeda, Inc. A GE Company

510(k) K152814

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

5 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK152814 Date17 Mar 2016ClearedStatusActive

Recalls

8 recalls · 5 ongoing · FDA

  • 2 recalls, same reason

    • Ongoing Z-2476-2026 FDA Class II initiated 1 May 2026
    • Ongoing Z-2477-2026 FDA Class II initiated 1 May 2026

    GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.

    • Ongoing Z-0048-2025 FDA Class I initiated 3 Sep 2024

    GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.

    • Ongoing Z-2100-2023 FDA Class II initiated 19 May 2023

    GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices

    • Ongoing Z-0281-2020 FDA Class II initiated 26 Jul 2019

    Fasteners potentially torqued to a value less than the specified value and can potentially cause the canopy to become detached and fall.

    • Completed Z-0140-2025 FDA Class I initiated 6 Jun 2023

    Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration.

    • Terminated Z-0539-2020 FDA Class I initiated 11 Oct 2019 on shared 510(k) K152814

    Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.

    • Terminated Z-0479-2019 FDA Class II initiated 25 Sep 2018

    Potential electric leakage affecting the user.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
FMZ
FDA applicant
Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company
Device Class
Class II
Regulation
21 CFR Part 880.5400
Advisory Committee
General Hospital
FDA Decision
SESE Substantially Equivalent

Overview

AI-generated summary, 1 Oct 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

The Giraffe OmniBed Carestation CS1 is an incubator, neonatal device designed to provide a controlled, protected microenvironment for neonatal care. It is intended for use in general hospital settings to support infant thermal regulation and healing.

The device is supplied as a reusable system and has received FDA 510(k) clearance under K152814. It is classified as a Class II medical device with product code FMZ and regulation number 880.5400.

Other devices from Ohmeda Medical, A Division of Datex-Ohmeda, Inc.

Questions

AI-generated answers, 1 Oct 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

  • What is the Giraffe OmniBed Carestation CS1 used for?

    The Giraffe OmniBed Carestation CS1 is designed to provide a controlled, protected microenvironment for neonatal care, supporting infant thermal regulation and healing in general hospital settings.

  • Is the Giraffe OmniBed Carestation CS1 reusable or single-use?

    The device is supplied as a reusable system, meaning it is intended for multiple uses after proper cleaning and reprocessing between patients.

  • What regulatory clearance does the Giraffe OmniBed Carestation CS1 have?

    The Giraffe OmniBed Carestation CS1 has received FDA 510(k) clearance under submission number K152814 and is classified as a Class II medical device with product code FMZ.