Global Blastocyst Vitrification Kit & Vitrification Thawing Kit - Based on S3
by <Genx> Intl., Inc. 510(k)
Active
510(k) K092963
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK092963 | Date14 Jan 2011 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-1421-2022
There was translation error in the Spanish version of the IFU where the Spanish version instructs the user to leave the blastocyte to air thaw for 5 minutes rather than the correct 5 seconds as indicated in the English version.
Specification
- 510(k) Number
- FDA Product Code
- MQL
- FDA applicant
- Genx Intl., Inc.
- Device Class
- Class II
- Regulation
- 21 CFR Part 884.6180
- Advisory Committee
- Obstetrics/Gynecology
- FDA Decision
- SESE Substantially Equivalent