GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) P42150
by Medtronic Sofamor Danek, Inc.
Listed
Valid until
Not applicable
FDA recalls
15 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00643169001060 | Date20 Jul 2015 | StatusListed |
Recalls
-
- Ongoing Z-0138-2027
Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured from the donor tissue. Donor bone tissue tested positive for Clostridium butyricum and while tissue did undergo gamma radiation, if implanted, there is a possibility of clostridium infection.
-
- Ongoing Z-0162-2024
- Ongoing Z-0163-2024
- Ongoing Z-0166-2024
- Ongoing Z-0165-2024
- Ongoing Z-0164-2024
- Ongoing Z-0167-2024
- Ongoing Z-0161-2024
Specific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with potential non-conformances in the outer Tyvek pouch.
-
- Ongoing Z-1325-2023
- Ongoing Z-1323-2023
- Ongoing Z-1326-2023
- Ongoing Z-1333-2023
- Ongoing Z-1328-2023
- Ongoing Z-1324-2023
- Ongoing Z-1327-2023
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
-
- Terminated Z-0977-2014
Product was labeled as being measured as 2.5 cm x 10 cm, but contained product that measured as 2.5 cm x 5 cm.
-
- Terminated Z-0084-2013
- Terminated Z-0087-2013
- Terminated Z-0083-2013
- Terminated Z-0088-2013
- Terminated Z-0091-2013
- Terminated Z-0085-2013
- Terminated Z-0086-2013
- Terminated Z-0094-2013
- Terminated Z-0092-2013
- Terminated Z-0090-2013
- Terminated Z-0089-2013
- Terminated Z-0093-2013
Multiple lots of GRAFTON and XPANSE , which are used for bone grafting procedures are being recalled because there is a possibility that sterility of the outer surface of the inner pouch may become compromised.
-
- Terminated Z-0415-2012
- Terminated Z-0416-2012
Aseptic conditions during the production of these products may have been compromised.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM)
- Generic Name
- Human-derived demineralized bone matrix implant
- Model / Reference
- P42150
- Primary DI / UDI-DI
- 00643169001060
- 510(k) Number
- FDA Product Code
- MBP, MQV
- GMDN Term
- Human-derived demineralized bone matrix implant
- Device Class
- Class II
- Regulation Number
- 888.3045
- GUDID distribution status
- Not in Commercial Distribution
- Size
- Length — 5.0 CentimeterWidth — 5.0 Centimeter
Other devices from Medtronic Sofamor Danek, Inc.
Properties
Other properties (8)
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- Premarket Exempt: No
- Serial Number: No
- Expiration Date: No
- Manufacturing Date: No