Greenlight Hps Series Surgical Laser System & Accessories
by Laserscope
510(k) K062719
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK062719 | Date1 Dec 2006 | StatusActive |
Recalls
-
- Ongoing Z-1392-2022
Several complaints were received for the affected lot; users received a "Fiber type does not match card type" error message on the GreenLight XPS Laser console, necessitating the replacement of the laser fiber and fiber card. This may result in prolonged procedure while the fiber is being replaced.
-
- Terminated Z-0700-2013
During electrical recertification testing it was found that the Voltage Scaling Board was non-conforming to the IEC 60601-1-2 standard.
-
- Terminated Z-2502-2010
Product may be packed with incorrect external or internal labels and/or missing product insert.
-
- Terminated Z-0131-2008
Control knob break: The knob used to finely control the direction of the laser energy may break.
-
- Terminated Z-0130-2008
Component missing: Some products were issued without thermal protection switches.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- GEX
- Device Class
- Class II
- Regulation
- 21 CFR Part 878.4810
- Advisory Committee
- General, Plastic Surgery
- FDA Decision
- SESE Substantially Equivalent