Greiner Vacuette Blood Collection Tube W/Lithium Heparin & Gel Separator
510(k) K960857
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK960857 | Date30 Aug 1996 | StatusActive |
Recalls
-
- Ongoing Z-0200-2027
Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.
-
- Terminated Z-0502-2022
Some of the blood collection tubes show clotting due to variation of Lithium Heparin.
-
- Terminated Z-1081-2012
The lithium heparin tubes may have a contamination with EDTA potassium.
-
- Terminated Z-0623-2011
Lithium heparin solution was contaminated by EDTA K (potassium) solution due to an operator error and a misinterpretation of a test result. Contaminated tubes will produce inaccurate test results.
-
- Terminated Z-0215-2009
Labeling Error: Incorrect tube label reads "Z Serum Clot Activator" instead of "LH Lithium Heparin" tube. All other information on the tube label including color coding of tube was correct. Case and rack label information showed the correct information with 2 ml Lithium Heparin.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- JKA
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1675
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent