  "format": "devicerts.device.v1",
  "cite_as": "Gridlock Plating System, Trilliant Surgical, LLC. deviCERTS medical device catalogue. https://devicerts.com/devices/gridlock-plating-system-hrs-k121452 (data as of 2026-11-21).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/gridlock-plating-system-hrs-k121452",
    "json_url": "https://devicerts.com/devices/gridlock-plating-system-hrs-k121452.json",
    "markdown_url": "https://devicerts.com/devices/gridlock-plating-system-hrs-k121452.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-11-08",
    "status_as_of": "2026-11-21",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/gridlock-plating-system-hrs-k121452",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Gridlock Plating System",
    "registry_name": "GRIDLOCK PLATING SYSTEM",
    "url": "https://devicerts.com/devices/gridlock-plating-system-hrs-k121452",
    "manufacturer": {
      "name": "Trilliant Surgical, LLC",
      "url": "https://devicerts.com/manufacturers/trilliant-surgical-llc",
      "country": null,
      "country_basis": null,
      "fda_registration_country": "United States"
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K121452",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "HRS"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K121452",
      "number_label": null,
      "device_udi_di": null,
      "date": "2012-08-20",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1609-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-07-19",
      "reason": "Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters.  This recall of IFU occurred in 2018.",
      "applies": "shared submission",
      "via": "K121452"
    },
    {
      "number": "Z-2490-2015",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-08-06",
      "reason": "The affected parts subject to the recall are out of specification, resulting in in the loss of the optimal locking screw functionality of the plates.  The thread depth of the screw holes is insufficient, potentially limiting the ability to engage and lock with associated mating locking screws.",
      "applies": "this device",
      "via": "K121452"
    },
    {
      "number": "Z-2738-2015",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-07-31",
      "reason": "Gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow engagement between the driver and the screw, resulting in an increased likelihood of stripping screw heads intraoperatively.  Specifically, the minor diameter of the drive feature is above the specification, potentially reducing the engagement between the driver and the screw head.",
      "applies": "this device",
      "via": "K121452"
    },
    {
      "number": "Z-2456-2015",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-07-20",
      "reason": "The cannulation of the Concave and Convex Reamer Heads may be too small to allow the Guide Pin to freely pass through.",
      "applies": "shared submission",
      "via": "K121452"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K121452",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "HRS",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 888.3030",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Orthopedic",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}