Griffin Spect/CT Imaging System
510(k) K041218
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK041218 | Date24 May 2004 | StatusActive |
Recalls
-
- Terminated Z-0513-2018
- Terminated Z-0514-2018
During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning against the cart), or it may fall/slide all the way to the floor.
-
- Terminated Z-1779-2015
- Terminated Z-1778-2015
The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
-
- Terminated Z-0256-2012
- Terminated Z-0258-2012
- Terminated Z-0257-2012
- Terminated Z-0255-2012
A mechanical malfunction of the button on the hand controller on the Philips Precedence and SKYLight SPECT systems may get stuck and cause unintended motion of a detector.
-
- Terminated Z-1550-2011
- Terminated Z-1549-2011
Philips Healthcare Nuclear Medicine determined that the detector arm assembly could be compromised.
-
- Terminated Z-0676-2008
Incorrect Measurement: The firm discovered a software defect in Cardiac and Pulmonary Viewer software application options used in the Precedence CT Sub-System. The defect causes incorrect measurements in derived images which are zoomed and saved in a batch file.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- KPS
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1200
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent