Guardian™ Connect App CSS7200
by Medtronic Minimed, Inc. 3 variants
Guardian Connect App CSS7200 is a medical device (FDA Class III; FDA product code MDS, Sensor, glucose, invasive) made by Medtronic Minimed, Inc., approved under FDA PMA P160007 on 8 Mar 2018 and listed in FDA GUDID since 2 Aug 2020 (UDI-DI 00763000381103), active as of 23 Oct 2026. 3 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00763000381103 | Date2 Aug 2020 | StatusListed |
| United States | FDA PMAAPPR | NumberP160007 | Date8 Mar 2018 | StatusActive |
Official records
- AccessGUDID record 00763000381103
- FDA PMA database P160007
From the FDA record
- Device description
- SFTWR CSS7200 GRDN MBL IOS V 4.1.0
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA PMA P160007 decided 8 Mar 2018 Active
- FDA GUDID UDI-DI 00763000381103 listed 2 Aug 2020 Listed
Recalls
-
- Ongoing Z-0033-2026
The device cap label may not adhere to the sensor, compromising the sterility of the sensor, and increasing the risk of infection.
-
- Ongoing Z-2106-2020
Customers using the firm's continuous glucose monitoring system application on an iPhone, iPad or iPod Touch with ios software version 12, 12.1, or 12,2 are likely to experience a shortened transmitter battery life (approximately 4-5 days instead of the normal 7 days or more) after a full charge.
-
- Terminated Z-1282-2019
The application may be closed by the operating system without alerting the user the app is no longer running or communicating with the transmitter resulting in the user not receiving alerts that could be associated with hypoglycemic or hyperglycemic events.
Timeline
- Approved FDA PMA P160007
- Recall initiated FDA recalls Z-1282-2019
- Recall initiated FDA recalls Z-2106-2020
- Listed FDA GUDID 00763000381103
- Recall initiated FDA recalls Z-0033-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA PMA Active
Specification
- Trade name
- Guardian™ Connect App
- Generic name
- Patient health record information system application software
- Model / reference
- CSS7200
- Primary DI / UDI-DI
- 00763000381103
- PMA number
- FDA product code
- MDS
- GMDN term
- Patient health record information system application software
- Device class
- Class III
- GUDID distribution status
- In Commercial Distribution
Similar devices in this category