  "format": "devicerts.device.v1",
  "cite_as": "GYN-Pump PH304, W.O.M. World of Medicine GmbH. deviCERTS medical device catalogue. https://devicerts.com/devices/gyn-pump-ph304-hig (data as of 2026-11-22).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/gyn-pump-ph304-hig",
    "json_url": "https://devicerts.com/devices/gyn-pump-ph304-hig.json",
    "markdown_url": "https://devicerts.com/devices/gyn-pump-ph304-hig.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-09",
    "status_as_of": "2026-11-22",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/gyn-pump-ph304-hig",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "GYN-Pump PH304",
    "registry_name": "GYN-Pump PH304",
    "url": "https://devicerts.com/devices/gyn-pump-ph304-hig",
    "manufacturer": {
      "name": "W.O.M. World of Medicine GmbH",
      "url": "https://devicerts.com/manufacturers/w-o-m-world-of-medicine-gmbh",
      "country": null,
      "country_basis": null,
      "fda_registration_country": "United States"
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K173489",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "HIG"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K173489",
      "number_label": null,
      "device_udi_di": null,
      "date": "2018-03-28",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-2561-2021",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2021-06-18",
      "reason": "When the display of inflow volume to the uterus reaches 32450 ml, the fluid management system control unit display will freeze because the internal software calculation threshold is reached. The outflow measurement will continue, and so the deficit accumulated up to this point will start counting backwards to 0 ml. There is a risk of distention fluid reaching the soft tissue circulatory system.",
      "applies": "this device",
      "via": "K173489"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K173489",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "HIG",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 884.1700",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Obstetrics/Gynecology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}