  "format": "devicerts.device.v1",
  "cite_as": "H-Scribe, Mortara Instrument, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/h-scribe-mlo (data as of 2026-11-19).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/h-scribe-mlo",
    "json_url": "https://devicerts.com/devices/h-scribe-mlo.json",
    "markdown_url": "https://devicerts.com/devices/h-scribe-mlo.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-10",
    "status_as_of": "2026-11-19",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/h-scribe-mlo",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "H-Scribe",
    "registry_name": "H-SCRIBE",
    "url": "https://devicerts.com/devices/h-scribe-mlo",
    "manufacturer": {
      "name": "Mortara Instrument, Inc.",
      "url": "https://devicerts.com/manufacturers/mortara-instrument-inc",
      "country": "United States",
      "country_basis": "FDA 510(k) applicant",
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K004017",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "MLO"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K004017",
      "number_label": null,
      "device_udi_di": null,
      "date": "2001-06-11",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0323-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-10-11",
      "reason": "There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),.  If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.",
      "applies": "this device",
      "via": "K004017"
    },
    {
      "number": "Z-0320-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-10-11",
      "reason": "There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),.  If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.",
      "applies": "this device",
      "via": "K004017"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K004017",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "MLO",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 870.2800",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Cardiovascular",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}