Habib EUS RFA
510(k) K161305
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK161305 | Date7 Jul 2016 | StatusActive |
Recalls
-
- Terminated Z-0668-2020
- Terminated Z-0667-2020
During testing on the Habib EUS RFA device, a temperature increase was observed at the tip of the EUS-FNA needle. It was determined that capacitive 5 coupling presents a risk of energy transfer from the Habib EUS RFA active tip to the introducer device. This capacitive coupling could result in thermal injury to tissue outside of the targeted treatment area due to its use with an un-insulated EUS-FNA needle requiring little or no medical intervention.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- GEI
- FDA applicant
- Emcision , Ltd.
- Device Class
- Class II
- Regulation
- 21 CFR Part 878.4400
- Advisory Committee
- General, Plastic Surgery
- FDA Decision
- SESE Substantially Equivalent