# Hamilton-C3

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/hamilton-c3-cbk>
- Manufacturer: [Hamilton Medical AG](https://devicerts.com/manufacturers/hamilton-medical-ag) (Switzerland)
- Name in the registry: HAMILTON-C3
- Device class: Class II (US FDA)
- FDA product code: CBK

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K161450 · decision 2017-04-27 · decision SESE · status: Active

## Recalls

- Z-1372-2018 · Class II · ongoing · initiated 2018-03-12: After two and a half years, the battery fuel gauge may indicate a higher battery charge than what is available.
- Z-2796-2018 · Class II · terminated · initiated 2018-03-09: An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer providing an acoustic alarm and the blinking alarm LED on the device s front side.

## Identifiers

- 510(k) Number: K161450 [FDA 510(k)]
- FDA product code: CBK [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 868.5895 [FDA 510(k)]
- Advisory committee: Anesthesiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Hamilton-C3, Hamilton Medical AG. deviCERTS medical device catalogue. https://devicerts.com/devices/hamilton-c3-cbk (data as of 2026-11-17).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-17
- JSON: <https://devicerts.com/devices/hamilton-c3-cbk.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/hamilton-c3-cbk
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.