Skip to content

Hamilton-G5

by Hamilton Medical AG 4 variants

510(k) K103803

Hamilton-G5 is a medical device (FDA Class II; FDA product code CBK, Ventilator, Continuous, Facility Use) made by Hamilton Medical AG, cleared under FDA 510(k) K103803 on 23 Nov 2011, active as of 22 Oct 2026. 4 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

4 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK103803 Date23 Nov 2011ClearedStatusActive

Official records

From the FDA record

Received by the FDA
28 Dec 2010
FDA decision
23 Nov 2011 (Substantially Equivalent)
FDA review time
330 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K103803 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

4 recalls · FDA

    • Terminated Z-1874-2019 FDA Class I initiated 25 Mar 2019

    New software version for affected ventilators reduces the probability of the ventilator entering an ambient state, in which the inspiratory channel and expiratory valves are opened, letting the patient breathe room air unassisted. When the ventilator enters the ambient state, alternative ventilation must be provided immediately.

    • Terminated Z-1624-2016 FDA Class II initiated 9 Mar 2016

    After performing the suctioning maneuver, including disconnecting the patient, suctioning , and reconnecting the patient, the preset pattern of ventilation many not continue as expected.

    • Terminated Z-2057-2015 FDA Class II initiated 24 Mar 2015

    Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and the user can no longer operate the device.

    • Terminated Z-1185-2012 FDA Not yet classified initiated 6 Feb 2012

    Hamilton-G5 ventilators with software version V2.0x may experience an unintentional change of the displayed ventilation mode under certain conditions.

Timeline

5 dated events

  1. Cleared FDA 510(k) K103803
  2. Recall initiated FDA recalls Z-1185-2012
  3. Recall initiated FDA recalls Z-2057-2015
  4. Recall initiated FDA recalls Z-1624-2016
  5. Recall initiated FDA recalls Z-1874-2019

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
CBK
Device class
Class II
Regulation
21 CFR Part 868.5895
Advisory committee
Anesthesiology
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code CBK (Ventilator, Continuous, Facility Use).

Other devices from Hamilton Medical AG

See all devices from Hamilton Medical AG →