  "format": "devicerts.device.v1",
  "cite_as": "Hamilton-G5, Hamilton Medical AG. deviCERTS medical device catalogue. https://devicerts.com/devices/hamilton-g5-cbk-k103803 (data as of 2026-11-27).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/hamilton-g5-cbk-k103803",
    "json_url": "https://devicerts.com/devices/hamilton-g5-cbk-k103803.json",
    "markdown_url": "https://devicerts.com/devices/hamilton-g5-cbk-k103803.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-27",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/hamilton-g5-cbk-k103803",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Hamilton-G5",
    "registry_name": "HAMILTON-G5",
    "url": "https://devicerts.com/devices/hamilton-g5-cbk-k103803",
    "manufacturer": {
      "name": "Hamilton Medical AG",
      "url": "https://devicerts.com/manufacturers/hamilton-medical-ag",
      "country": "Switzerland",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K103803",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "CBK"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K103803",
      "number_label": null,
      "device_udi_di": null,
      "date": "2011-11-23",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1874-2019",
      "classification": "Class I",
      "status": "terminated",
      "initiated": "2019-03-25",
      "reason": "New software version for affected ventilators reduces the probability of the ventilator entering an ambient state, in which the inspiratory channel and expiratory valves are opened, letting the patient breathe room air unassisted. When the ventilator enters the ambient state, alternative ventilation must be provided immediately.",
      "applies": "this device",
      "via": "K103803"
    },
    {
      "number": "Z-1624-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-03-09",
      "reason": "After performing the suctioning maneuver, including disconnecting the patient, suctioning , and reconnecting the patient, the preset pattern of ventilation many not continue as expected.",
      "applies": "this device",
      "via": "K103803"
    },
    {
      "number": "Z-2057-2015",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-03-24",
      "reason": "Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and the user can no longer operate the device.",
      "applies": "this device",
      "via": "K103803"
    },
    {
      "number": "Z-1185-2012",
      "classification": "",
      "status": "terminated",
      "initiated": "2012-02-06",
      "reason": "Hamilton-G5 ventilators with software version V2.0x may experience an unintentional change of the displayed ventilation mode under certain conditions.",
      "applies": "this device",
      "via": "K103803"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K103803",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "CBK",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 868.5895",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Anesthesiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}