Heater Cooler Unit, Model Hcu 30
by Jostra AB 510(k)
Active
510(k) K031544
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK031544 | Date7 Aug 2003 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-2140-2021
- Terminated Z-0348-2017
There is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.
-
- Terminated Z-0589-2016
Failure of the HEATER-COOLER UNIT (HCU 30) Type 2 power supply board. A drop of more than 10% voltage can cause overheating and subsequent failing of the R56 thermistor of the power supply board. Board failure typically occurs when the compressor is switched on (prior to patient use); however, it can occur during patient use. This can shut down the compressor or the entire HCU 30 Type 2 unit.
-
- Terminated Z-0350-2007
The HCU 30 system potentially will not be have the power capacity to run the compressor when the heaters are on during operation. This may lead to a situation where the user may not be able to cool the patient and/or cardioplegia sufficiently.
Specification
- 510(k) Number
- FDA Product Code
- DWC
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.4250
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent