Hemoglobin A1c Assay, Hemoglobin A1c Calibrators, Hemoglobin A1c Controls
by Abbott Laboratories 510(k)
Active
510(k) K140654
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK140654 | Date22 May 2015 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-1454-2022
There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using the ARCHITECT and Alinity c Hemoglobin A1c assay.
-
- Terminated Z-1688-2020
Calibrator level 2 bottle may have been manufactured incorrectly for a portion of the lot. The manufacturer observed no impact to assay performance between the HbA1c concentration range of 5.33% to 6.87%. However, outside of this range, a bias of greater than 3% may be seen with patient results. The device failure that is associated with this recall can result in potentially negatively and positively biased patient test results for HbA1c. The firm is investigating the root cause.
Specification
- 510(k) Number
- FDA Product Code
- PDJ
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1373
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent