Hemotherm Model 400
510(k) K811742
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK811742 | Date13 Aug 1981 | StatusActive |
Recalls
-
- Terminated Z-2720-2011
- Terminated Z-2721-2011
The probe jack contacts (terminals) on certain of the firm's ECMO Heater devices are located too close to the solid state relay contacts (terminals) which can lead to the terminals coming into contact with each other. If this situation occurs, the service technician, user, and/or patient could be exposed to electric shock during servicing and/or use.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DWJ
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.5900
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent