Skip to content

Hemotherm Model 400CE Cooler/Heater System

by Cincinnati Sub-Zero Products, Inc.

510(k) K122813

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

6 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 6 Oct 2026. What this means

Regulatory status

as of registry sync 6 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK122813 Date18 Dec 2012ClearedStatusActive

Recalls

6 recalls · FDA

    • Terminated Z-2092-2021 FDA Class II initiated 17 Jun 2021

    There is a potential risk of device contamination and patient infection associated with the device because of the potential for organisms (including Nontuberculous mycobacteria NTM) to grow in the water systems of any heater-cooler device, and contaminated water from any heater-cooler device has the potential to aerosolize into the operating room during surgery which could lead to patient infection.

    • Terminated Z-0825-2019 FDA Class II initiated 19 Dec 2018

    Evidence supports that when customers performed a fuse replacement as part of a March 2017 field action, the fuse clips became stretched and were no longer mechanically tight with the fuse, creating a host spot that led to discoloration on the board due to overheating and shut down of the device.

    • Terminated Z-0063-2019 FDA Class II initiated 4 Sep 2018

    Potential for membrane (control panel) switch errors leading to an inability of the device to change control settings or an unintended change to control settings.

    • Terminated Z-0629-2018 FDA Class II initiated 27 Nov 2017

    Due to FDA's safety communication relating to nontuberculous mycobacteria (NTM) infections related to heater/cooler devices, an updated Instructions For Use includes a more robust cleaning & disinfection procedure.

    • Terminated Z-1825-2017 FDA Class II initiated 9 Mar 2017

    Improperly rated fuses were used in the manufacture of the device.

    • Terminated Z-0654-2017 FDA Class II initiated 7 Nov 2016 on shared 510(k) K122813

    The device may not maintain water flow or temperature control.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
DWC
Device Class
Class II
Regulation
21 CFR Part 870.4250
Advisory Committee
Cardiovascular
FDA Decision
SESE Substantially Equivalent

Overview

AI-generated summary, 28 Sep 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

The Hemotherm Model 400CE Cooler/Heater System is a controller for temperature management in cardiopulmonary bypass procedures. It is designed to provide precise blood temperature control during cardiovascular surgeries and related procedures.

The device is supplied as a reusable system for use in clinical settings such as operating rooms. It requires cleaning and maintenance between uses and is not intended for single-use disposal. The system operates under FDA 510(k) clearance (K122813) as a Class II medical device with product code DWC, regulated under 870.4250 for temperature controllers in cardiopulmonary bypass applications.

Other devices from Cincinnati Sub-Zero Products, Inc.

See all devices from Cincinnati Sub-Zero Products, Inc. →

Questions

AI-generated answers, 28 Sep 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

  • What is the Hemotherm Model 400CE Cooler/Heater System used for?

    The Hemotherm Model 400CE Cooler/Heater System is a controller designed to provide precise blood temperature control during cardiopulmonary bypass procedures, such as cardiovascular surgeries and related procedures, by managing the temperature of blood circulated through the bypass circuit.

  • Is the Hemotherm Model 400CE Cooler/Heater System intended for single-use or reuse?

    The Hemotherm Model 400CE Cooler/Heater System is supplied as a reusable system for use in clinical settings such as operating rooms. It requires cleaning and maintenance between uses and is not intended for single-use disposal, ensuring it can be safely reprocessed for multiple procedures.

  • What regulatory pathway was used for the Hemotherm Model 400CE Cooler/Heater System?

    The Hemotherm Model 400CE Cooler/Heater System operates under FDA 510(k) clearance (K122813) as a Class II medical device with product code DWC, regulated under 870.4250 for temperature controllers in cardiopulmonary bypass applications, based on a determination of substantial equivalence to a predicate device.