Skip to content

Hispeed Dual

by GE Medical Systems, Inc.

510(k) K021491

Hispeed Dual is a medical device (FDA Class II; FDA product code JAK, System, X-Ray, Tomography, Computed) made by GE Medical Systems, Inc., cleared under FDA 510(k) K021491 on 22 May 2002, active as of 22 Oct 2026. One FDA recall is linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

2 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK021491 Date22 May 2002ClearedStatusActive

Official records

From the FDA record

Received by the FDA
9 May 2002
FDA decision
22 May 2002 (Substantially Equivalent)
FDA review time
13 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K021491 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

2 recalls · FDA

    • Terminated Z-2224-2009 FDA Unclassified (pre-2012) initiated 1 Dec 2008

    During internal testing, GE Healthcare has recently determined that the collimator used on our CT/e and HiSpeed X/i families of Computed Tomography systems does not comply with the specific requirements in United States Title 21 of the Code of Federal Regulations, and IEC 60601-1-3, related to leakage radiation from the diagnostic source assembly, which includes the collimator. The excess leak

    • Terminated Z-0269-05 FDA Unclassified (pre-2012) initiated 21 Oct 2004 on shared 510(k) K021491

    The patient tables of the affected CT systems are supported by a hydraulic cylinder and a cylinder shaft (i.e. the upper shaft) which connects the tabletop to the table frame. GE received a report of a HiSpeed LX/I system table falling abruptly due to the filure of the upper shaft. No injuries were reported.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

3 dated events

  1. Cleared FDA 510(k) K021491
  2. Recall initiated FDA recalls Z-0269-05
  3. Recall initiated FDA recalls Z-2224-2009

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
JAK
Device class
Class II
Regulation
21 CFR Part 892.1750
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code JAK (System, X-Ray, Tomography, Computed).

Other devices from GE Medical Systems, Inc.

See all devices from GE Medical Systems, Inc. →